length, or failure to connect cable shields properly, may also expose the unit
to electrical noise, resulting in an unstable meter reading.
Figure 5-3: Cable ends
B
C
A
A. Unshielded length
B. Coil
C. Electrode
WARNING
Shock hazard! Potential shock hazard across remote junction box terminals 1
and 2 (85V).
WARNING
Explosion hazard! Electrodes exposed to process. Use only compatible
transmitter and approved installation practices. For process temperatures
greater than 284°F (140°C), use a wire rated for 257°F (125°C).