14.8 Europe
14.8.1 I1 ATEX Intrinsic Safety
Certificate
Baseefa09ATEX0125X
Standards
IEC 60079-0:2011; EN60079-11:2012;
Markings
II 1G Ex ia IIC T4 Ga, T4(–50 °C≤ T
a
≤ +70 °C)
Special Conditions for Safe Use (X):
1. The surface resistivity of the antenna is greater than 1GΩ. To avoid
electrostatic charge build-up, it must not be rubbed or cleaned with
solvents or dry cloth.
2. The Rosemount 775 enclosure may be made of aluminum alloy and
given a protective polyurethane paint finish; however, care should be
taken to protect it from impact or abrasion if located in zone 0.
14.8.2 N1 ATEX Type n
Certificate
Baseefa09ATEX0131
Standards
IEC 60079-0:2012 + A11:2013, EN 60079-15:2010;
Markings
II 3G Ex nA IIC T4 Gc, T4(–50 °C≤ T
a
≤ +70 °C) IP66
14.9 International
14.9.1 I7 IECEx Intrinsic Safety
Certificate
IECEx BAS 09.0050X
Standards
IEC 60079-0:2011, IEC 60079-11:2011
Markings
Ex ia IIC T4 Ga, T4(–50 °C≤ T
a
≤ +70 °C) IP66
Special Conditions for Safe Use (X):
1. The surface resistivity of the antenna is greater than 1 GΩ. To avoid
electrostatic charge build-up, it must not be rubbed or cleaned with
solvents or dry cloth.
2. The Rosemount 775 enclosure may be made of aluminum alloy and
given a protective polyurethane paint finish; however, care should be
taken to protect it from impact or abrasion if located in zone 0.
14.9.2 N7 IECEx Type n
Certificate
IECEx BAS 09.0058
Standards
IEC 60079-0:2011, IEC 60079-15:2010;