EMC ENVIRONMENT
Guidance and Manufacturer's declaration - electromagnetic emissions
EmbracePlus is intended for use in the electromagnetic environment
specified below. The user of the EmbracePlus should assure that it is
used in such an environment.
Emission Test
Compliance
Electromagnetic environment
RF emission
CISPR 11
Group 1
EmbracePlus uses RF energy only for
its internal function. Therefore, its
RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
RF emission
CISPR 11
Class B
The EmbracePlus device is suitable
for use in domestic establishments
and in establishments directly
connected to the low voltage power
supply network that supplies
buildings used for domestic
purposes.
Harmonic
Emission
EN 61000-3-2
Not
Applicable
Voltage
Fluctuation/
flicker
emission
EN 61000-3-3
Not
Applicable