EasyManua.ls Logo

Empatica EmbracePlus - Page 128

Empatica EmbracePlus
133 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1. This device may not cause harmful interference, and
2. This device must accept any interference received, including
interference that may cause undesired operation (FCC Title 47,
Subpart A, Part 15.19(3)).
Changes or modifications not expressly approved by the party
responsible for compliance could void the user’s authority to operate
the equipment (FCC Title 47, Subpart A, Part 15.21)
GUIDANCE AND MANUFACTURER'S DECLARATION -
ELECTROMAGNETIC IMMUNITY
The EmbracePlus is intended for use in the electromagnetic
environment specified below. The user of the EmbracePlus should
assure that it is used in such an environment.
Immunity Test
IEC 60601
test level
Compliance
Level
Electromagnetic environment-
guidance
Electrostatic
Discharge
(ESD) EN
61000-4-2
8 kV
contact
2/4/8/15
kV Air
IEC
60601-1-2
Test level
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30 %
Radiated
electromagn
etic field
IEC
61000-4-3
10 V/m
80 MHz to
2.7 GHz
IEC
60601-1-2
Test level
Portable and mobile RF
communications
equipment should be used
no closer to any part of
EmbracePlus, including
127