continued
Adjustable
Rate
Options:
The
following
program
options
(9-12)
offer
adjustable
rate.
To
change
rate,
press
both
PROG+
and
PROG-
buttons
at
the
same
time
while
device
is
on
and
set
on
one
of
the
rate
adjustable
program
options
(9,
10,
11,
or
12).
Display
will
show
the
program
letter
(C,
B,
R,
or
M)
and
the
current
rate.
Use
PROG+
or
PROG-
to
go
to
the
next
or
previous
rate.
Press
PROG+
and
PROG-
buttons
at
the
same
time
when
selection
is
complete.
Automatic
return
to
program
operation
will
occur
after
one
minute.
9
Continuous
(C)
100
pps
vere
Pe
This
setting
produces
a
continuous
stimulation
at
the
set
intensity
and
a
rate
of
100
pps.
10
Cycled
Burst
(B)
In
the
Cycled
Burst
program,
stimulation
is
on
for
2
seconds
and
off
for
2
seconds.
11
Rate
Modulation
(R)
100
pps
brbehe+-beb
oe
In
the
Modulated
Rate
program,
for
0.5
second
the
rate
is
at
100
pps
and
for
the
next
0.5
second,
it
is
at
50
pps.
The
cycle
is
then
repeated.
There
is
no
off
time.
Both
channels
are
used
simultaneously.
12
Multi-modulation
(M)
100
pps
In
the
Multi-modulated
program,
for
0.5
second
the
rate
is
at
100
pps and
the
intensity
is
at
60%
of
set
level
and
for
the
next
0.5
second,
the
rate
is
at
50
pps
while
the
intensity
is
at
100%
of
set
level.
The
cycle
is
then
repeated.
There
is
no
off
time.
Both
channels
are
used
simultaneously.
Data
Retrieval
Controls
The
Epix
VT
device
stores
usage
and
pain
scale
data.
This
feature
is
for
clinician
use
only.