117
No water jet coming from the Piezon
handpiece
 
 Check the water adjustment of your unit
 
 Check for presence of water at the cord outlet port
 Send the handpiece to an authorized EMS Repair Center


 
 
 Eliminate residual powder present in the channels using the cleaning needles provided
 Check the quantity of powder in your unit
 Send the system to an authorized EMS Repair Center
Water leakage between the handpiece
cord and the handpiece
 
 Send the system and the handpiece cord to an authorized EMS Repair Center

handpiece and the nozzle
 Verify the position of the nozzle and its position on the handpiece
 Verify that no dirt remains between the nozzle and the handpiece
 
 Replace the nozzle
 Replace the handpiece
 
Air and powder leakage on the main unit  Verify the positioning of the bottles
 
 
 
 Remove all powder dust between the bottle holder and the powder chamber
 Send the unit to an authorized EMS Repair Center
Air and powder is leaking at powder
chamber
 Check the seal condition and the cleanliness of the thread of the powder chamber cap
 Replace the seals if necessary
 Clean the sealing surfaces carefully
 Replace the powder chambers
 Send the unit to an authorized EMS Repair Center


 Verify the quantity of powder in your unit
 Eliminate residual powder present in the channels using a cleaning needle
 
Leakage on the unit water supply  Check the water supply connection and inlet pressure