Table of Contents

Questions and Answers

  • S
    summer79Aug 18, 2025
    What to do if no water jet comes from the EMS Dental equipment handpiece?
    • M
      Morgan LeeAug 18, 2025
      If no water jet is coming from the handpiece of your EMS Dental equipment when the footpedal is activated: * Ensure the unit is switched on and check all power cable connections and the dental office's power supply. * Check the filter and water adjustment settings on your unit. * Verify the unit's water supply and the office's water supply. * Inspect the connections between the unit, cord, and handpiece. * Check for water presence at the connector cord output (without the handpiece) and at the unit connector cord (without the cord). * If the water channel of the cord is blocked, blow through it with compressed air (max 3 bar). If blockage persists, or if water is present at the cord outlet but not the handpiece, send the affected part to an approved EMS repair center.
  • A
    Ashley JacksonAug 20, 2025
    How to fix control lamp not illuminating on EMS AIR-FLOW S2 Dental equipment?
    • R
      Rodney PhillipsAug 20, 2025
      If the control lamp on your EMS Dental equipment does not illuminate when you switch it on: * Ensure the unit is switched on using the power switch. * Check the connections of the unit's power cable. * Verify the power supply of the dental office. * If the problem persists, send the unit to an approved EMS repair center.
  • A
    Anthony PerezAug 25, 2025
    Why is my EMS AIR-FLOW S2 Dental equipment leaking water between the handpiece cord and handpiece?
    • H
      heather42Aug 25, 2025
      If you notice water leaking between the handpiece cord and the handpiece, inspect the O-ring on the handpiece cord. If the problem persists, send the handpiece cord to an approved EMS repair center.
  • R
    Rachael CampbellAug 27, 2025
    What to do if there are no ultrasonic oscillations when activating the footpedal on my EMS Dental equipment?
    • A
      Ann HarrellAug 27, 2025
      If you're not getting any ultrasonic oscillations when you press the footpedal, try these steps: * Ensure the unit is powered on. * Check that the airflow handpiece cord is properly placed in its holder. * Verify the unit's air connection. * Inspect all connections between the unit, the handpiece cord, and the handpiece itself. * Check the footpedal connection. * Try changing the handpiece. If none of these steps work, the unit may require servicing at an approved EMS repair center.
  • J
    Jeanette BrownAug 31, 2025
    Why is my EMS AIR-FLOW S2 Dental equipment leaking on the unit water supply?
    • D
      Debra FoleyAug 31, 2025
      If there's a leak on the unit's water supply, first check the water supply connection. If that doesn't solve the problem, try changing the water supply hose with a connector. If the leakage continues, send the unit to an approved EMS repair center.
  • T
    Travis SolomonSep 3, 2025
    How to fix air and/or powder leaking from the cap of the powder chamber of EMS Dental equipment?
    • M
      Miranda GarciaSep 3, 2025
      If air or powder is leaking from the threads of the powder chamber cap, check the seal and cleanliness of the threads on both the powder chamber and the cap. Replace the seal if needed. Clean the cap and powder chamber. If the problem continues, change the cap. If all else fails, send the unit to an approved EMS repair center.
  • W
    William GarzaSep 7, 2025
    Why is the ultrasonic power decreasing or insufficient in my EMS Dental equipment?
    • S
      Stacey RasmussenSep 7, 2025
      If you're experiencing decreasing or insufficient ultrasonic power, ensure the instrument is properly positioned. Check the instrument for wear and replace it if necessary. You could also try changing the handpiece. If the issue persists, send the handpiece to an approved EMS repair center.
  • N
    Nancy BrownSep 10, 2025
    Why control lamp do not illuminate when the EMS Dental equipment is switched on?
    • J
      Joseph BurnettSep 10, 2025
      If the control lamp doesn't illuminate when you switch on the unit, start by ensuring the unit is switched on using the power switch. Next, check the connections of the power cable to the unit and verify the power supply of the dental office. If the lamp still doesn't light up, send the unit to an approved EMS repair center.
  • M
    matthew70Nov 13, 2025
    What to do if no powder/air jet comes from the EMS AIR-FLOW S2 handpiece?
    • S
      Sandra NguyenNov 13, 2025
      If there is no powder or air coming from the handpiece of your EMS Dental equipment when you activate the footpedal, try the following: * Ensure the unit is switched on. * Check all power cable connections to the unit. * Verify the dental office's power supply. * Make sure the Piezon handpiece cord is properly placed in its holder. * Check the unit's air connection and all connections between the unit, cord, and handpiece. * Inspect the footpedal connection and the compressed air supply. * Clean and dry the powder chamber; check powder/air presence at the cord outlet (without handpiece) and the system connector outlet (without cord). * Eliminate any residual powder in the handpiece using cleaning needles. * Check the powder quantity in the unit. If the issue persists, co...

Summary

Welcome and Initial Safety

Product Introduction and Overview

Welcome message and overview of the EMS Air-Flow® S2.

Essential Pre-Operation Warnings

Critical instructions and safety precautions to read before operating the unit.

Installation Procedures

Device Components

Identification of all parts and control elements of the EMS Air-Flow® S2 unit.

Connect Footpedal

Step-by-step guide for connecting the footpedal to the unit.

Connect Water Supply

Guide on connecting the unit to the water supply.

Connect Compressed Air Supply

Instructions for connecting the unit to the compressed air supply.

Connect Electrical Supply

Guide on connecting the unit to the electrical power supply.

Operating the Unit

Basic Unit Operation

Overview of unit controls and how to switch it on/off.

AIR-FLOW Mode Operation

Using the AIR-FLOW® function, including flow rate and powder settings.

PIEZON Mode Operation

Using the PIEZON® function, including water flow and ultrasonic power settings.

Maintenance and Troubleshooting

Cleaning, Disinfecting, and Sterilizing

Procedures for cleaning, disinfecting, and sterilizing unit parts.

Routine Maintenance Procedures

Regular maintenance tasks for components like O-rings and powder chambers.

Troubleshooting Guide

Solutions for common problems and malfunctions of the unit.

EMS AIR-FLOW S2 Specifications

General IconGeneral
Powder Flow RateAdjustable
TypeAir Polisher
ApplicationDental prophylaxis
Water SupplyExternal water supply
TechnologyAIR-FLOW®
Particle Size40 µm
Nozzle TypesPerio and standard
Powder TypeEMS AIR-FLOW powder
Water Flow RateAdjustable