70
TEHNIČ KI PODACI
OPIS
Proizvođač EMS SA, CH-1260 Nyon, Švicarska
Model
Air-Flow
®
S2
KlasikacijaEN60601-1 - Klasa I
- Primjenjiv dio, Tipa B
- IP 20 aparat
-IPX1,nožnapoluga
Klasikacija93/42CEE Klasa IIa
Načinrada
sa ispiranjem:
bez ispiranja:
Kontinuirano djelovanje
Diskontinuirano 10% kroz maksimalno
10 minuta
Napon napajanja 100 -110 VAC / 50 / 60 Hz
120 - 130 VAC/ 50 / 60 Hz
220 VAC/ 50 / 60 Hz
230 - 240 VAC / 50 / 60 Hz
Potrošnja max.80VA
Osigurači
cijevni 5 x 20 mm, 250 VAC, časové
zpoždění,nízkývypínacívýkon
za 100 – 130 VAC napajanje: T1AL
za 220 – 240 VAC napajanje: T0.5AL
Napajanje vodom 1 do 5 bara (100 do 500 kPa)
Napajanje komprimiranim zrakom 4,5 do 7 bara (450 do 700 kPa)
Ultrazvučnespecikacije
Maksimalna snaga:
Opseg frekvencije:
8 W
25 - 32 kHz
Težina Približno4.0kg
Dimenzijeumm(visinaxširinax
dubina)
120x250x250
Uvjeti funkcioniranja +10°C do +40°C
30%do75%relativnevlažnosti
Maximálnínadmořskávýška3000m
Uvjetičuvanja
i transporta
-10°C do +40°C
10%do95%relativnevlažnosti
500 hPa do 1060 hPa tlaka zraka
TECHNICAL DATA
DESCRIPTION
Manufacturer EMS SA, CH-1260 Nyon, Switzerland
Model
Air-Flow
®
S2
ClassicationEN60601-1 - Class I
- Applied part, Type B
- IP 20 unit
- IP X1, footpedal
Classication93/42EWG Class IIa
Mode
with irrigation:
without irrigation:
Continuous operation
Dutycycle10%during10min.maximum
Supply voltage 100 -110 VAC / 50 / 60 Hz
120 - 130 VAC/ 50 / 60 Hz
220 VAC/ 50 / 60 Hz
230 - 240 VAC / 50 / 60 Hz
Power consumption max.80VA
Fuses
Ø5x20mm,250VAC,timelag,lowbreak-
ing capacity
for 100 - 130 VAC power supply: T1AL
for 220 - 240 VAC power supply: T0.5AL
Water supply 1 to 5 bar (100 to 500 kPa)
Compressed air supply 4.5 to 7 bar (450 to 700 kPa)
Ultrasonicspecications
Maximumoutput:
Frequency range:
8 W
25 - 32 kHz
Weight Approx.4.0kg
Dimensionsinmm(HxWxD) 120x250x250
Operating conditions +10° C to +40° C
30% to 75% relative humidity
Altitudemax.3000m
Storage and
transport conditions
-10°C to +40°C
10% to 95% relative humidity
500 hPa to 1060 hPa air pressure