TROUBLESHOOTING
Symptom Solutions
The display does not light up when
the unit is switched on
s Check the different electrical connections
s Check the mains
s Check/replace the fuses and replace them only with the type indicated on the back of the unit
s Send the unit to your authorized EMS repair center
Handpiece without pressure or with
a low pressure when the foot switch
is activated
s Check the different air connections
s Check the pressure available for the unit
s Check that the manometer is indicating more than 1.5 bar
s Check the compressor
s Check the connection between the handpiece and unit
s Test the unit with another handpiece
s Send the unit to your authorized EMS repair center
Pressure not adjustable s Check the different air connections
s Check the pressure available for the unit
s Check that the manometer is indicating more than 1.5 bar
s Check the compressor
s Send the unit to your authorized EMS repair center
Pressure reduction during operation s Check the different air connections
s Check the compressed air supply
s Contact your authorized EMS repair center
Handpiece without pulse when the
foot switch is actived
s Check that the "Validate" indicator light is on
s Check the connection of the foot pedal
s Check the different connections of the unit
s Check the different supplies of the unit
s Test the unit with another handpiece
Pressure too high s Send the unit to your authorized EMS repair center