Installation Liquistation CSF33
14 Endress+Hauser
4.1.4 Connection for suctioning samples
Maximum suction height:
Vacuum pump: Standard 6 m (20 ft)
Peristaltic pump: standard 8 m (26 ft)
Maximum hose length: 30 m (98 ft)
Hose connection diameter
Vacuum pump: 13 mm (1/2")
Peristaltic pump: internal diameter of 10 mm (3/8")
Intake speed:
> 0.6 m/s (> 1.9 ft/s) for 10 mm (3/8") ID, as per Ö 5893, US EPA
> 0.5 m/s (> 1.6 ft/s) for ≤ 13 mm (1/2") ID, in accordance with EN 25667, ISO 5667
Note the following when erecting the device:
Always route the suction line so that it slopes upwards from the sampling point to the
sampler.
The sampler must be located above the sampling point.
Avoid siphoning effects in the suction line.
Requirements for the sampling point:
Do not connect the suction line to pressurized systems.
Use the suction filter to impede coarse and abrasive solids and solids which can cause
clogging.
Immerse the suction line in the direction of flow.
Take the sample at a representative point (turbulent flow, not directly at the bottom of the
channel).
Useful sampling accessories
Suction filter:
Impedes coarser solids and solids which can cause clogging.
4.1.5 Connection for sample intake on version with pump
Maximum suction height:
Vacuum pump: Standard 6 m (20 ft)
Peristaltic pump: standard 8 m (26 ft)
Maximum hose length: 30 m (98 ft)
Hose connection diameter
Vacuum pump: 13 mm (1/2")
Peristaltic pump: internal diameter of 10 mm (3/8")
Intake speed:
> 0.6 m/s (> 1.9 ft/s) for 10 mm (3/8") ID, as per Ö 5893, US EPA
> 0.5 m/s (> 1.6 ft/s) for ≤ 13 mm (1/2") ID, in accordance with EN 25667, ISO 5667
Note the following when erecting the device:
Always route the suction line so that it slopes upwards from the sampling point to the
sampler.
The sampler must be located above the sampling point.
Avoid siphoning effects in the suction line.