Liquistation CSF33 Installation
Endress+Hauser 15
Requirements for the sampling point:
Do not connect the suction line to pressurized systems.
Use the suction filter to impede coarse and abrasive solids and solids which can cause
clogging.
Immerse the suction line in the direction of flow.
Take the sample at a representative point (turbulent flow, not directly at the bottom of the
channel).
Useful sampling accessories
Suction filter:
Impedes coarser solids and solids which can cause clogging.