Calculations Tankvision
32 Endress+Hauser
Typical table example: 54B (1980)
L00-NXA82xxx-16-00-00-xx-046
Special corrections
VCF Chemical 1
L00-NXA82xxx-16-00-00-xx-049
with:
•D
Ref
= Reference density
•T
Obs
= Actual or observed temperature
•T
Ref
= Reference Temperature
• A-E = Configurable coefficients
TCF Method
L00-NXA82xxx-16-00-00-xx-050
with:
•K
0
, K
1
, K
2
, K
3
, K
4
= Customer coefficients
• t = Actual product temperature
•T
Ref
= Reference Temperature
For the VCF we can write:
L00-NXA82xxx-16-00-00-xx-051
With the TCF method the p
Ref
can be calculated with p
Obs
/ VCF (when manual p
Obs
is
provided) and an actual p
Obs
can be calculated with p
Ref
x VCF, when p
Ref
is known and
no actual p
Obs
is available.
The range of application:
Density, kg/m³ Temperature, °C
653.0 to 778.5 -18.0 to 90.0
779.0 to 824.0 -18.0 to 125.0
824.5 to 1075.0 -18.0 to 150.0