EasyManua.ls Logo

EngA DJM3 - CALIBRATION; DISCHARGE SENSOR; DISCHARGE SETPOINT CALIBRATION; ROOM SENSOR CALIBRATION

EngA DJM3
27 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A DJM3
IOM-17 23 of 27 Mar 99 R3
INDUCEDVOLTAGEONREMOTECONTROLWIRING
TheDJM3controliscalibratedatthefactoryandthiscalibrationwillsatisfymostapplications.
Occasionallyimpedancefromlongwiringrunsorinducedvoltageontheremotewiringcauseittobe
outofcalibration.Torecalibraterefertothefollowingsection.Tocheckforinducedvoltages,
disconnectallofthefieldwiringinquestionfromtheDJM3.2whiletheunitisleftoperating,andcheck
forACvoltsacrosseachquestionablewireandground.
CALIBRATION
DischargeSensor
1. Accuratelymeasurethedischargeairtemperatureatthetemperaturesensorin°C.
2. ReadtheDJM3dischargetemperaturebyreadingthevoltageatmultimetertestpointDTEMP.
UsingaDCvoltmeter
3. IfadifferenceisfoundadjusttheDSENCALPottomatchtothetwovalues
DischargeSetpointCalibration
1. RecordthesettingsofDIPswitches1,2,and3.TurnoffDIPswitches1,2,and3.Ifthe“OR”
terminalisused,thewiretoterminal“OR”mustberemoved.
2. Dependingonwhichsetpointcontrolmethodisused:
a) JohnsonControlsTE6100
i) SetJohnsonControlsTE6100setpointto21°C
ii) ReadtheDJM3DischargesetpointatmultimetertestpointSPCusingaDCvoltmeter.
iii) AdjusttheDISSPCALPOTtomatchthereading.
b) DJM3BuiltinDial
i) Turnsetpointdialto21°C.
ii) ReadtheDJM3DischargesetpointatmultimetertestpointSPCusingaDCvoltmeter.
iii) Loosentheset‐screwonthesetpointknobandmovetheknobmarkertomatchwiththe
measuredvalue.
iv) Tightenset‐screw.
Foreitheroftheabove:
3. ReturntheDIPswitchestotheirnormalposition.
RoomSensorCalibration
1. Accuratelymeasuretheroomorreturnairin°Candsettheroomsetpointdialtothemeasured
value.
2. RecordthesettingsofDIPswitches1,2,and3.TurnoffDIPswitches1,2,and3.
3. ReadtheDJM3DischargesetpointatmultimetertestpointSPCusingaDCvoltmeter.
4. TurnDIPswitch3on
5. ReadtheDJM3DischargesetpointagainandadjusttheROOMCALPottomatchthesamevoltage
measuredinstep3.
6. ReturntheDIPswitchestotheirnormalposition.

Related product manuals