During treatment, the patient may not feel unpleasant sensations amounting to pain. A mild
sensation of excitation is permissible.
If, as result of treatment, headache, vertigo, fatigue and/or other (autonomic nervous)
reactions develop, subsequent treatment should be given at a lower intensity.
Before treatment
Please make sure you have read and understood the content of this manual before you start
treatment.
Check the patient for any possible contra-indications.
The skin of the area in question is cleaned (removal of grease) with soap or 70 % alcohol. Shaving
hairy skin is recommended.
Test the sensitivity of the treatment area.
Position the electrodes and/or sponges (do not forget to moisten).
During treatment
Intensity is set at the desired level.
The patient is regularly asked to report any sensations felt. If necessary, the treatment is modified.
Ending treatment
Treatment can be stopped by pressing the stop icon on the touchscreen or setting the timer to zero.
Treatment stops automatically when the treatment time has elapsed.
After treatment
Remove the electrodes and/or sponges.
Clean the patient’s skin with a towel or tissue.
The expected effects are checked (e.g. pain, circulation, and mobility).
The patient is asked to comment subsequently on any reactions that may occur.