Operators Manual
Combination
Blanket/Fluid Warmer
EC1350BL
EC1850BL
MN-39940
REV.03
8/20
EN

Questions and Answers

  • G
    Gwendolyn PalmerAug 18, 2025
    What to do if the sensor is shorted in Enthermics EC1350BL Medical Equipment?
    • L
      Lauren GallegosAug 18, 2025
      To address a shorted sensor in your Enthermics Medical Equipment, begin by testing the sensor. Detach it from the warmer and use an Ohm meter to measure its resistance at 25°C (77°F). If the reading is 10 KOhm ±1.5 KOhm, replace the display. If it's ±2 KOhm, replace the sensor. Also, check the wires for any damage and ensure all connections at the control and terminal block are secure. Resecure any faulty connections. If the problem continues, it is recommended to call for service.
  • A
    Aaron HernandezAug 20, 2025
    What to do if the sensor is open in Enthermics Medical Equipment?
    • C
      Christopher SmithAug 20, 2025
      To troubleshoot an open sensor in your Enthermics Medical Equipment, start by testing the sensor. Detach it from the warmer and use an Ohm meter to measure the resistance at 25°C (77°F). If the reading is 10 KOhm ±1.5 KOhm, replace the display. If it's ±2 KOhm, replace the sensor. Check the wires for any damage and ensure secure connections at the control and terminal block; re-secure if necessary. If the issue remains, contact service.
  • A
    abrownAug 23, 2025
    What to do if Enthermics EC1350BL Medical Equipment shows heater pad over-temp error?
    • L
      Lorraine JuarezAug 23, 2025
      If you encounter a heater pad over-temp error on your Enthermics Medical Equipment, acknowledge the error by pressing the ON/OFF button. Allow the warmer to cool down. If the error continues, contact service.
  • F
    Frederick JohnsonAug 26, 2025
    How to cool down Enthermics EC1350BL if the cavity temperature is too high?
    • J
      Jillian ColeAug 26, 2025
      If the cavity temperature of your Enthermics Medical Equipment is higher than the set point, allow the cavity to cool to the set point temperature.
  • P
    pricesaraAug 27, 2025
    What to do if the temperature difference between cavity sensors 1 and 2 is too high in Enthermics EC1350BL?
    • N
      ndelacruzAug 28, 2025
      If the temperature difference between cavity sensors 1 and 2 exceeds 1.7ºC (3ºF) in your Enthermics Medical Equipment, remove the product and allow the warmer to cool down. Ensure the product sensor is clean and functioning correctly. Press the power button to clear the error code. If the error persists, contact service to replace the sensor switch assembly.
  • P
    Paula DawsonAug 30, 2025
    How to fix real time clock battery discharge in Enthermics EC1350BL?
    • T
      Timothy McbrideAug 30, 2025
      If the real-time clock rechargeable battery backup has discharged in your Enthermics Medical Equipment, plug the warmer into the outlet for 30 minutes. Then, refer to the Timer Control Feature section in the blanket manual to reset the clock.
  • M
    Michael CaldwellSep 1, 2025
    What to do for Analog to Digital Convertor Error on Enthermics EC1350BL Medical Equipment?
    • K
      Kevin ThompsonSep 2, 2025
      If you get an Analog to Digital Convertor Error on your Enthermics Medical Equipment, remove any inventory and discard it if necessary. Allow the warmer to cool down. If the error persists after cooling and resetting, contact service to replace the control assembly.
  • M
    Michelle GarciaSep 4, 2025
    How to adjust the number of heater pads on Enthermics Medical Equipment?
    • T
      Terri WaltersSep 5, 2025
      If your Enthermics Medical Equipment detects more heater pads than set, hold the ON/OFF button for 12 seconds until the display shows "PAd#" (# = number of pads selected [3-7]). Use the up or down arrow to adjust to the correct number of pads.
  • J
    Joshua BellSep 7, 2025
    What happens if a button is held down too long on Enthermics Medical Equipment?
    • T
      Tyler TaylorSep 7, 2025
      If a button has been held down for more than 60 seconds on your Enthermics Medical Equipment, adjust the control. The error will reset once the problem is resolved.
  • N
    Nicole WagnerSep 9, 2025
    What to do if Enthermics Medical Equipment displays a Hardware Over Temp error?
    • J
      jweaverSep 9, 2025
      If you are experiencing a Hardware Over Temp error with your Enthermics Medical Equipment, inspect the connections and condition of the high limit bimetal thermostat. If the error persists, contact service.

Enthermics EC1350BL Specifications

General IconGeneral
CategoryMedical Equipment
ModelEC1350BL
MaterialStainless Steel
Power120V
FeaturesDigital temperature control, audible alarm

Related product manuals