Table 16: long term calibration plan
Calibration
UVB frames
VIS frames
NIR
frames
Frequency
Purpose
DARK_UVB_100k
3x1hour
monthly
dark
DARK_UVB_400k
3x1hour
monthly
dark
DARK_VIS_100k
3x1hour
monthly
dark
DARK_VIS_400k
3x1hour
monthly
dark
DARK_UVB_100k_1x2
3x1hour
monthly
dark
DARK_UVB_400k_1x2
3x1hour
monthly
dark
DARK_VIS_100k_1x2
3x1hour
monthly
dark
DARK_VIS_400k_1x2
3x1hour
monthly
dark
DARK_UVB_100k_2x2
3x1hour
2 months
dark
DARK_UVB_400k_2x2
3x1hour
2 months
dark
DARK_VIS_100k_2x2
3x1hour
2 months
dark
DARK_VIS_400k_2x2
3x1hour
2 months
dark
Long darks NIR
3x1hour
On
request
dark
LINEARITY_UVB_100k
Set of detector
FF + biases
monthly
detector monitoring
LINEARITY_UVB_400k
Set of detector
FF + biases
monthly
detector monitoring
LINEARITY_VIS_100k
Set of detector
FF + biases
monthly
detector monitoring
LINEARITY_VIS_400k
Set of detector
FF + biases
monthly
detector monitoring
LINEARITY_UVB_100k_1x2
Set of detector
FF + biases
monthly
detector monitoring
LINEARITY_UVB_400k_1x2
Set of detector
FF + biases
monthly
detector monitoring
LINEARITY_VIS_100k_1x2
Set of detector
FF + biases
monthly
detector monitoring
LINEARITY_VIS_400k_1x2
Set of detector
FF + biases
monthly
detector monitoring
LINEARITY_UVB_100k_2x2
Set of detector
FF + biases
monthly
detector monitoring
LINEARITY_UVB_400k_2x2
Set of detector
FF + biases
monthly
detector monitoring
LINEARITY_VIS_100k_2x2
Set of detector
FF + biases
monthly
detector monitoring
LINEARITY_VIS_400k_2x2
Set of detector
FF + biases
monthly
detector monitoring
LINEARITY_NIR
Set of
detector
FF
monthly
detector monitoring
Imaging mode
darks
Skyflats
linearity
monthly
All of these calibrations are taken for the monitoring of the instrument health but also for
calibrating the science and calibration (telluric and flux standard stars) observations. One
should use at the time of the data reduction both daily and long-term calibrations in the
different corresponding pipeline recipes.