EasyManua.ls Logo

Epelsa ECO MULTIFUNCTION - Autozero Functionality

Epelsa ECO MULTIFUNCTION
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GBS-ECO-MF-V02
20
take several seconds if the weight is not stable at that time. During the time that the
scale takes to zero, the reading of the zero's value in internal points is displayed.
The scale then automatically sets a known weight value and shows the following
display:
At this time, the weight value detected by the scale is shown, even if it is an
incorrect value because it has not been adjusted (in this case '0.000', there is no
weight on the platform).
Place the known weight on the platform and enter the value of the known
standard weight (2/3 of the full-scale value is recommended). After entering the
value, press the key. It is possible to correct the weight by pressing the
key. The scale displays the reference values and the reference zero for
several seconds, which it will save as references of the setting time. It will then
show the real value of the weight set:
To return to the basic work mode, press the calibration button again. The
scale's zero and weight will now be set.
0.000
WE ADJ
0.000
Max 15 kg
Min 100 g
e= 5 g
I I I
0
kg
kg
Pcs
T
PT
10.000
WE ADJ
10.000
Max 15 kg
Min 100 g
e= 5 g
I I I
0
kg
kg
Pcs
T
PT