EasyManua.ls Logo

E+E Elektronik EE771 Series - 5.4.2. 2-point adjustment

E+E Elektronik EE771 Series
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
10% 20% 30% 40% 50% 60% 70% 80% 90% 100%
100%
90%
80%
70%
60%
50%
40%
30%
20%
10%
upper adjustment point at 80% of measuring range
lower adjustment point automatically at 0% of m.r.
characteristic line before adjustment
selected adjustment point
characteristic line after adjustment
measuring range
output signal
10% 20% 30% 40% 50% 60% 70% 80% 90% 100%
100%
90%
80%
70%
60%
50%
40%
30%
20%
10%
lower adjustment point at 20% of measuring range
upper adjustment point automatically at 100% of m.r.
selected adjustment point
characteristic line before adjustment
measuring range
output signal
characteristic line after adjustment
5.4.2. 2-point adjustment
With a 2-point adjustment procedure the lower adjustment point must be between 0 and 40% of the measuring
range, and the upper adjustment point between 60 and 100% of the measuring range. If the adjustment point is
between 40 and 60% of the measuring range, automatically a 1-point adjustment procedure will be executed instead.
lower adjustment point upper adjustment point
possibility 1 0 - <40% of m.r. 60 - 100% of m.r.
possibility 2 40 - <50% of m.r. 100% of m.r.
possibility 3 0% of m.r. 50 - <60% of m.r.
10% 20% 30% 40% 50% 60% 70% 80% 90% 100%
100%
90%
80%
70%
60%
50%
40%
30%
20%
10%
lower adjustment point at 10% of measuring range
upper adjustment point at 90% of measuring range
characteristic line before adjustment
characteristic line atfer adjustment
measuring range
output signal

Table of Contents