4. Maintenance
4.1 Cleaning
The stand-aid lifter may be cleaned using a slightly damp cloth with a standard household
cleaning agent. Chemicals and other strong/abrasive agents must not be used.
The stand aid lifter cannot under any circumstance be autoclaved.
It is recommended that the stand-aid lifter be cleaned after each use to avoid infection. The
stand-aid lifter can be disinfected with household alcohol.
4.2 Storage
The stand-aid lifter must be stored in a dry environment with relative humidity under 90%.
If the stand-aid lifter is not to be used over a longer period of time it is best to fully charge
the unit and activate the emergency button to save battery before putting in storage.
4.3 Special environments
The stand-aid lifter is not intended for use in aggressive environments e.g. leisure centres or
swimming pools. For these situations please contact your local dealer or Ergolet.
5. Check and trouble shooting
5.1 Annual check / inspection
The stand-aid lifter must always be maintained by qualified and authorised personnel who
have been approved by Ergolet.
The manufacturer Ergolet stipulates that this lifter MUST be checked by a qualified and
authorised person at least once a year. This annual check is also recommended in the
Standard (EN/ISO 10535) for lifters.
It is the customer’s / dealer’s responsibility to ensure that the annual service inspection is
carried out.
If it is necessary to replace any parts due to wear or damage these must be
obtained/purchased from Ergolet or your local dealer.