Operation Manual
Version: 12/01
25
er900 PC Operation Manual
Art-Nr: 475.156
Ergometer
Type er900 PC
Annex A: Technical Data
Annex A: Technical Data
Type : ergometrics 900 PC
Operation mode: continuous operation
Power supply: 115 V / 230 V ~/ 50 Hz / 60 Hz
Power rating: 90 VA
Braking principal: computer controlled eddy current brakes with torque regula-
tion (RPM independent in accordance with DIN 13405)
Load range: 20 . . . 999 Watt
RPM range: 30 . . . 130 n/min
-1
Deviation of load : acc. to DIN 13405: max. ± 5% in range 25 and 400 W
however: max. ± 3 W in range 25 and 100 Watt
acc. to manufacturer: max. ± 5% in range 20 and 900 W
however: max. ± 3 W in range 20 and 100 W
Load increments: manual setting: 5W programmable: 5W or multiple thereof
Max. admissable patient weight: 120 kg
Seat adjustment: continuously adjustable for body sizes 1.20m to 2.10m
Handle bar adjustment: continuously adjustable for body sizes 1.20m to 2.10m
Display: 1x LCD-Display : 4 lines @ 20 char.
1 x LED 10 mm: rpm display for patient
Blood pressure measurement : RR indirect with automatic artefact suppression
Pulse measurement : priority principal: 1. ECG, 2. Opto, 3. RR
measurement range: 35 to 230 bpm
Electrical inputs: Digital control: RS232
Analog: Opto, ECG 1V
Electrical outputs:
Digital: RS232,
Analog: load, pulse, systole, diastole
remote start ECG (1 to 99 sec)
Weight: 64 kg
Dimensions (W x L): min.: 400 x 900 mm, width of handle bar: approx. 540 mm
height: min: 950 mm; max: 1330 mm
Calibration: performed with switches and keypad
Protection of calibration: service plug
Safety standard: BF according to IEC 601-1
Device class: I according to IEC 601
Classification MDD: class IIa according to 93/42 ECC