Replacement frequency of filtration breakthrough sensor,
Molecode option
Molecular lters saturation detection alarm – Molecode (S : solvents or A : acids or F: formaldehyde - tted as an integrated
option in your unit, work using semi-conductor or electrochemical sensor that must be changed every :
5 years for the Molecode S
2 years for Molecode A and F
After this period, Erlab cannot guarantee performances stability nor detection sensitivity.
Equipment replacement requires the intervention of a qualied engineer, able to replace internal components and perform
system new conguration. The change requires less than one hour.
Erlab Maintenance department and accredited partners can provide this service. (service not available in all countries)
For further information, please get in touch with your contact person or Erlab dealer/distributor or visit our website
Admissible weighs on working surfaces
Tempered glass
work top
Stainless steel
work top
Trespa
®
Top Lab
PLUS
work top
50 kg/m
2
1187 lb/ft
2
110 kg/m
2
2610 lb/ft
2
110 kg/m
2
2610 lb/ft
2
Captair Smart 321 15 kg 33 lbs 35 kg 77 lbs 35 kg 77 lbs
Captair Smart 391 20 kg 44 lbs 45 kg 99 lbs 50 kg 110 lbs
Captair Smart 481 25 kg 55 lbs 65 kg 143 lbs 65 kg 143 lbs
Captair Smart 392 20 kg 44 lbs 60 kg 132 lbs 60 kg 132 lbs
Captair Smart 483 30 kg 66 lbs 80 kg 176 lbs 80 kg 176 lbs
Captair Smart 633 40 kg 88 lbs 100 kg 220 lbs 100 kg 220 lbs
Captair Smart 714 45 kg 99 lbs 115 kg 254 lbs 115 kg 254 lbs