EasyManua.ls Logo

EST Centurion W - Page 77

EST Centurion W
205 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Figure 4oo: Vial 91 Calibration Screen
21. The L/R axis is the only axis that needs adjustment for the Vial 91 position. (Note:
If there are adjustments made to the F/R axis, they will not be saved.) If this axis
needs adjustment, refer again to step 10. When the position is calibrated, <click>
Update Calibration.
22. This concludes the water axis calibration. If you have a Centurion W, the calibration
routine will conclude at this point. If you have a Centurion WS, the soil axis also
needs to be calibrated, proceed to 23.
23. This step begins with the vial gripped in tray position 1.
Figure 4pp: Vial 1/Tray Position Pop-up
24. <click> OK and the Vial 1/Tray position can be calibrated, see Figure 4qq.

Related product manuals