Press .
Press
or to select the desired value (15.0 to 30.0).
Press
to confirm.
9.11 3.7 ACAL (Auto Conductivity Calibration)—Con & TDS only
See Section 5.1 for more details on Automatic Conductivity Calibration.
Press .
Press
or to select “Yes” (Automatic) or “No” (Manual).
Press
to confirm.
9.12 3.8 SPC (Single Point Calibration)—Con & TDS only
See Section 5.2 for details on Single Point & Multi Point Calibration.
Press .
Press
or to select “Yes” (Single) or “No” (Multi).
Press
to confirm.
9.13 3.8 CELL (Nominal Cell Constant)—Con & TDS only
The PC 700 includes a probe with a nominal cell constant (k) of 1.0.
Use probes with k = 0.1 and 10 (sold separately) for improved
performance in extreme sample ranges. Use this setup function to
change the cell constant if necessary. Meter default is 1.0 to match the
included probe.
k = 0.1 ideal for low measurements <20 µS (<10 ppm).
k = 1.0 ideal for mid-range measurements
k = 10 ideal for high measurements >20 mS (>10 ppt).
Press .
Press
or to select 0.1, 1.0, or 10.0
Press to confirm.
15