pH Range | -2.00 to 16.00 pH |
---|---|
Calibration Points | Up to 5 points with Auto-buffer recognition, From 2 to 5 consecutive points; 0.1, 1.0, 10.0, 100.0 or 1000.0 ppm |
Buffer Options | USA : pH 1.68, 4.01, 7.00, 10.01, 12.45 NIST: pH 1.68, 4.01, 6.86, 9.18, 12.45 |
mV Range | ± 2000 mV |
Temperature Range | 0.0 to 100.0 ºC / 32.0 to 212.0 ºF |
Calibration | Offset in 0.1 º increments; Offset range: ± 5.0 ºC / 9.0 ºF |
Ion Range (Ion 700 only) | 0.01 to 2000 ppm ( ± 2000mV) |
Resolution | 0.01 pH, 0.1 mV (± 199.9 mV), 1 mV beyond ± 200 mV, 0.1 ºC / ºF |
Accuracy | ± 0.01 pH, ± 0.2 mV (± 199.9 mV), ± 2 mV beyond ± 200 mV, ± 0.3 ºC/± 0.5 ºF (0 to 70 ºC), ±0.5% FS (mono-valent) ±1.0% Full Scale (di-valent) |
Offset Adjustment | Up to ±150 mV |
Compensation | Automatic or Manual (0 to 100 ºC / 32 to 212 ºF) |
Resolution (ppm) | 0.01/ 0.1/1 ppm |
Slope Range | 15 mV/decade to 90 mV/decade |
Slope Display | Yes (with offset) |
Memory | 100 data sets |
Inputs | BNC, ATC, Reference (Half-cell) |
Power | AC/DC 9V, 6W Adapter (100/240 VAC, 50-60Hz) |
Guides on calibrating the meter for accurate pH measurements using standard buffers.
Outlines the procedure for calibrating the temperature sensor for accurate readings.
Explains how to adjust the mV offset for relative ORP measurements.
Details the calibration process for Ion Selective Electrodes (ISE) on the Ion 700 meter.