6. TROUBLESHOOTING GUIDE
PROBLEM CAUSE SOLUTION
No display a) Main power not
switched on.
b) AC Adapter
socket not
inserted
properly.
a) Switch on the power supply.
b) Re-insert AC Adapter & press
.
a) Insufficient
reference
electrolyte in
electrode.
a) Fill electrode with reference electrolyte.
b) Broken
electrode.
b) Replace electrode.
c) External ‘noises’
or induction
(e.g. electrical
‘noise’ caused
by a nearby
motor).
c) Remove or switch off interfering device.
Unstable
reading
d) Dirty electrode. d) Clean & rehydrate the electrode.
Slow
response
a) Dirty electrode. a) Clean & rehydrate the electrode.
Meter not
responding
to key press
a) HOLD mode
in operation.
b) Bad keypad.
a) Press to deactivate. See SETUP
to disable Auto Hold feature if enabled.
b) Contact Technical Service.
11

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EUTECH INSTRUMENTS Ion 700 and is the answer not in the manual?

EUTECH INSTRUMENTS Ion 700 Specifications

General IconGeneral
pH Range-2.00 to 16.00 pH
Calibration PointsUp to 5 points with Auto-buffer recognition, From 2 to 5 consecutive points; 0.1, 1.0, 10.0, 100.0 or 1000.0 ppm
Buffer OptionsUSA : pH 1.68, 4.01, 7.00, 10.01, 12.45 NIST: pH 1.68, 4.01, 6.86, 9.18, 12.45
mV Range± 2000 mV
Temperature Range0.0 to 100.0 ºC / 32.0 to 212.0 ºF
CalibrationOffset in 0.1 º increments; Offset range: ± 5.0 ºC / 9.0 ºF
Ion Range (Ion 700 only)0.01 to 2000 ppm ( ± 2000mV)
Resolution0.01 pH, 0.1 mV (± 199.9 mV), 1 mV beyond ± 200 mV, 0.1 ºC / ºF
Accuracy± 0.01 pH, ± 0.2 mV (± 199.9 mV), ± 2 mV beyond ± 200 mV, ± 0.3 ºC/± 0.5 ºF (0 to 70 ºC), ±0.5% FS (mono-valent) ±1.0% Full Scale (di-valent)
Offset AdjustmentUp to ±150 mV
CompensationAutomatic or Manual (0 to 100 ºC / 32 to 212 ºF)
Resolution (ppm)0.01/ 0.1/1 ppm
Slope Range15 mV/decade to 90 mV/decade
Slope DisplayYes (with offset)
Memory100 data sets
InputsBNC, ATC, Reference (Half-cell)
PowerAC/DC 9V, 6W Adapter (100/240 VAC, 50-60Hz)