PROBLEM CAUSE SOLUTION
From any screen
Invalid key;
Button does not
work in the current
operation mode.
Press alternate key.
Select valid buttons
depending on mode.
Ion CAL mode Ion slope calibration error;
calibration solutions are
outside acceptable range
(15 to 90 mV per decade).
Recalibrate using fresh ion
solutions. Use required ISA.
Replace electrode.
Ion CAL mode Calibration points
are more than 1
decade apart.
Recalibrate using
consecutive calibration
values 1 decade apart
(ex. 1.0 and 10.0).
Ion CAL mode Calibration attempted
with only 1-point during
ion calibration.
Recalibrate using 2 or more
ion calibration points.
pH CAL mode Calibration attempted
using only 1-point that
was not pH 7 (USA) or
pH 6.86 (NIST).
Recalibrate using 2 or more
pH points or use pH 7 (USA)
or use pH 6.86 (NIST).
Err with electrode
icon
Calibration error.
Buffer value does not
match value displayed or
electrode is disconnected
or failing.
Use fresh buffer solutions.
Check electrode connection.
Clean & recondition
electrode.
Replace electrode.
Ur (Under range)
Or (Over range)
Measured value is out
of range.
Electrodes not connected.
Electrode clogged, dirty
or broken.
Meter not calibrated.
Wrong temp value.
Make sure electrode is
connected.
Clean or replace electrode.
Recalibrate the meter.
Treat samples to bring
within meter measuring
range.
Reset meter.
12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EUTECH INSTRUMENTS Ion 700 and is the answer not in the manual?

EUTECH INSTRUMENTS Ion 700 Specifications

General IconGeneral
pH Range-2.00 to 16.00 pH
Calibration PointsUp to 5 points with Auto-buffer recognition, From 2 to 5 consecutive points; 0.1, 1.0, 10.0, 100.0 or 1000.0 ppm
Buffer OptionsUSA : pH 1.68, 4.01, 7.00, 10.01, 12.45 NIST: pH 1.68, 4.01, 6.86, 9.18, 12.45
mV Range± 2000 mV
Temperature Range0.0 to 100.0 ºC / 32.0 to 212.0 ºF
CalibrationOffset in 0.1 º increments; Offset range: ± 5.0 ºC / 9.0 ºF
Ion Range (Ion 700 only)0.01 to 2000 ppm ( ± 2000mV)
Resolution0.01 pH, 0.1 mV (± 199.9 mV), 1 mV beyond ± 200 mV, 0.1 ºC / ºF
Accuracy± 0.01 pH, ± 0.2 mV (± 199.9 mV), ± 2 mV beyond ± 200 mV, ± 0.3 ºC/± 0.5 ºF (0 to 70 ºC), ±0.5% FS (mono-valent) ±1.0% Full Scale (di-valent)
Offset AdjustmentUp to ±150 mV
CompensationAutomatic or Manual (0 to 100 ºC / 32 to 212 ºF)
Resolution (ppm)0.01/ 0.1/1 ppm
Slope Range15 mV/decade to 90 mV/decade
Slope DisplayYes (with offset)
Memory100 data sets
InputsBNC, ATC, Reference (Half-cell)
PowerAC/DC 9V, 6W Adapter (100/240 VAC, 50-60Hz)

Related product manuals