Rev. 041311-BT
%T
90
80
70
60
50
40
30
20
10
0
987654 3 2 1 0
14
CH
TR7
eXact
®
Strip Micro CH, Part No. 481657-II - for 4mL Samples
Sodium Chloride results require the table below. Follow eXact
®
Micro 7+ Chloride (as NaCl) Test Procedure (above)
using
eXact
®
Strip Micro CH, Part No. 481657-II. NOTE: To convert the NaCl value to Chloride (Cl
–
),
multiply the value from the chart below by 0.6. (Example: 1380ppm NaCl = 828ppm Cl
–
)
Chloride
(as NaCl)
Table
7
Find the “TR7” result in the table below to determine the Sodium Chloride concentration in ppm
(parts per million). (Example: a “TR7” result of 65.3 (use only the 65 for the chart) equals a Sodium
Chloride value of 1400 ppm). Record result.
After testing is completed, rinse cell immediately.
USE TABLE
6
Dip the eXact
®
Strip Micro CH (as NaCl), Part No. 481657-II into the CELL and immediately
press READ. This starts the 20 SECOND countdown timer. During this time move the strip
in a gentle back and forth motion. Remove and discard the strip after “1” on the display
disappears*. The cursor will move across the display while the meter prepares to measure
the sample. Record result displayed (this result is automatically stored in TR7).
DIP STRIP
5
Press the ZERO/ON button.
The cursor will move across the display,
followed by 100 %T.
Meter is ready for testing.
ZERO METER*
4
Use the small pipet provided to add 0.2ml of the sample to the water in the cell.
This is a
1 to 20 dilution of the pool water that is required to perform this test. NOTE: If you are using the
more accurate dilution procedure with the Dilution Kit (487200) or Mini Dilution Kit II (487202),
follow the procedure as provided with those kits.
ADD THE SAMPLE
3
Using deionized or distilled water, rinse the CELL at least 3 times to minimize potential cross
FRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOWKHFHOOWRWKHWRSZLWKGHLRQL]HGRUGLVWLOOHGZDWHU
PREPARE THE CELL
2
Press and re-press the MENU button until the display shows the parameter TR7.
SELECT TEST: TR7
1
Press the ZERO/ON button to power the meter on; the display will show all annunciators,
then the current MENU selection, followed by the last reading.
TURN METER ON
Salt/Chloride (as NaCl) Test Procedure
For Pool and Spa Salt Systems
0 40 80 120 160 200 240 280 320 360
400 440 480 520 560 600 640 680 720 760
800 840 880 960 1000 1040 1080 1120 1160 1200
1240 1280 1320 1360 1400 1440 1480 1520 1560 1600
1640 1680 1720 1760 1800 1840 1880 1920 1960 2000
2040 2060 2100 2140 2180 2200 2240 2280 2320 2360
2400 2460 2500 2540 2580 2620 2660 2700 2740 2780
2820 2860 2900 2940 2980 3040 3080 3140 3200 3240
3280 3340 3400 3460 3520 3580 3640 3700 3760 3840
3940 4040 4200 4280 4440 4660 5000 5840 7000 >7000
*NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.