Bromine DPD-1 Test Procedure
1
Remove one (1) eXact
®
Strip Micro CL (DPD-1), Part No. 486637 from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
6
Dip the eXact
®
Strip Micro CL (DPD-1), Part No. 486637 into the CELL and immediately press
READ. This starts the 20 SECOND countdown timer. During this time move the strip in a gentle
back and forth motion (approx. 2 strokes/sec). Remove and discard the strip after “1” on the
display disappears. The cursor will move across the display while the meter prepares to
measure the sample. Record result displayed (this result is stored in BR3 memory). After
testing is completed, rinse cell immediately and press MENU to select next test.
DIP STRIP AND PRESS “READ”
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
The sample is ready for testing.
ZERO METER
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
TURN METER ON
3
Press and re-press the MENU button until the display shows the parameter BR3.
SELECT TEST: BR3
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\POZLWK
the water sample.
FILL METER WITH SAMPLE
1
Remove one (1) eXact
®
Strip Micro AL, Part No. 486641 from the bottle before
beginning the test. Set the strip in a dry, convenient place and recap the bottle immediately.
REMOVE STRIP
6
Dip the eXact
®
Strip Micro AL, Part No. 486641 into the CELL and immediately press READ. This
starts the 20 SECOND countdown timer. During this time move the strip in a gentle back and forth
motion (approx. 2 strokes/sec). Remove and discard the strip after “1” on the display disappears*.
NOTE: For water temperatures above 95°F/35°C (hot tubs), remove and discard the strip when the
timer displays “10”, countdown continues. For the hot water samples, a 10-second dip time is best.
The cursor will move across the display while the meter prepares to measure the sample. Record
UHVXOWGLVSOD\HGWKLVUHVXOWLVDXWRPDWLFDOO\VWRUHGLQ$/$IWHUWHVWLQJLVFRPSOHWHGULQVHFHOO
immediately.
*NOTE: When testing outdoors (sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading the sample.
DIP STRIP AND PRESS “READ”
5
Press the ZERO/ON button. The cursor will move across the display, followed by 0.00 PPM.
The sample is ready for testing.
ZERO METER*
2
Press the ZERO/ON button to power the meter on; the display will show all annunciators, then
the current MENU selection, followed by the last reading.
TURN METER ON
3
Press and re-press the MENUEXWWRQXQWLOWKHGLVSOD\VKRZVWKHSDUDPHWHU$/
SELECT TEST: AL4
4
Rinse the CELL at least 3 times with the water sample you will be testing - rinsing minimizes
WKHSRWHQWLDOIRUFURVVFRQWDPLQDWLRQIURPDSUHYLRXVWHVW)LQDOO\¿OOFHOOWRFDSDFLW\POZLWK
the water sample.
FILL METER WITH SAMPLE
Total Alkalinity Test Procedure
CL
BR3
AL
AL4