8
C+RMODE:Capacitance is shown on the Primary Display and the
parallel resistance R, is shown on the Secondary Display. The units
of capacitance are pF, nF, or µF. Resistance units are or k. Parallel
(C < 500 pF) or serial (C > 500 pF) equivalent circuit diagram is used.
C+DMODE:Capacitance is shown on the Primary Display and
dissipation factor D on the Secondary Display. The capacitance
is either the equivalent series or parallel capacitance of the
DUT. The units of capacitance are pF, nF, μF or mF. D is the
ratio of the real part of the impedance to the imaginary part
of the impedance, or 1/Q. D is dimensionless and the same
for series and parallel representations. A good capacitor has
a large C (imaginary) and a small R (real) and thus a low D.
|Z|MODE:The Impedance of the component is shown
on Primary Display. Units are mΩ, , k, or MΩ.
ESRMODE:The equivalent series resistance of the capacitor is
shown on the Primary Display. ESR units are mΩ, , k, or MΩ.
AUTOMODE:IT determines which component model is the most
accurate representation of the DUT and automatically selects the
appropriate parameter set. The determination is made as follows:
For |Q| < 0.15 the R mode is selected.
For Q > +0.15 the L+R or L+Q mode is selected (depends on user settings).
For Q < -0.15 the C+R or C+D mode is selected.
For C < 500 pF Parallel circuit diagram (Rp) is used.
For C >= 500 pF Serial circuit diagram (Rs) is used.
MENUSTRUCTURESANDFUNCTIONS
This section describes menu structure and device parameters
setting. IntelliTweeze menu system contains
Main menu — main menu items
System menu — system menu items
Sound menu – sound settings
Display menu — display settings