23
APPENDIXA.SPECIFICATIONS
TECHNICALSPECIFICATIONS
AC test mode Test
frequency:
1 kHz, 10 kHz, 120Hz,100 Hz
Test frequency accuracy: 50 PPM (0.005%)
Test signal level: 0.5/1.0 +/- 5% Vrms Sine wave
Source impedance: 100Ω +/- 1%
TYPICALOFFSET:
Resistance ≤ 25 mΩ
Capacitance 0.65 pF
Inductance 100 uH
Offset value should be subtracted from measurement result for
small value components (R < 10Ω , C < 100 pF, L < 10 μH).
MEASUREMENTRANGES
Parameter Measurement Range Test frequency
Resistance < 9.9 MΩ 1 kHz
Capacitance < 9999 pF 10 kHz
10000 pF to 10 μF 1 kHz
> 10 uF 100 Hz
Inductance 0.5 μH to 1 mH 10 kHz
1 mH to 1000 mH 1 kHz
> 1000 mH 100 Hz