P
IN
mode
23
For example, after the first reading the display may show:
This means that sample number
1
has a value of 3.27x10
-3
SI
Units. Remember that the display is always in 10
-3
SI Units.
Therefore, a value of 0.01 = 0.01x10
-3
SI Units = 1x10
-5
SI Units,
which is the limit of sensitivity for the KT-9.
Additional readings
1
To take additional measurements, the above steps
1
to
5
are repeated.
2
Each new measurement is automatically stored in memory
when you press the
P
IN
for the next reading. The digit after
the C: will change to reflect the current sample number.
3
Up to
10
samples can be stored in memory. When the
eleventh sample is taken, the display will show:
The XXX represents the value of the reading and the
F indicates that the memory is now full.
4
If you continue taking additional readings after the tenth,
the eleventh sample is moved into memory location ten.
All the other samples are moved down one memory loca-
tion, resulting in the first sample being lost.
1:3.27
If you do not wish to store a particular measurement,
then you must press the L
EFT
button before taking the next
reading.
F:XXX
P
IN
mode
23
For example, after the first reading the display may show:
This means that sample number
1
has a value of 3.27x10
-3
SI
Units. Remember that the display is always in 10
-3
SI Units.
Therefore, a value of 0.01 = 0.01x10
-3
SI Units = 1x10
-5
SI Units,
which is the limit of sensitivity for the KT-9.
Additional readings
1
To take additional measurements, the above steps
1
to
5
are repeated.
2
Each new measurement is automatically stored in memory
when you press the
P
IN
for the next reading. The digit after
the C: will change to reflect the current sample number.
3
Up to
10
samples can be stored in memory. When the
eleventh sample is taken, the display will show:
The XXX represents the value of the reading and the
F indicates that the memory is now full.
4
If you continue taking additional readings after the tenth,
the eleventh sample is moved into memory location ten.
All the other samples are moved down one memory loca-
tion, resulting in the first sample being lost.
1:3.27
If you do not wish to store a particular measurement,
then you must press the L
EFT
button before taking the next
reading.
F:XXX