14 TM500-EU-EN v1.1 12/13
ThermocoupleInput
(typesK,J,T,E,R,andS)
Thesuppliedthermocoupleprobesareuseableintherang
eof‐20to+250°C(4to+482°F)only.For
measuringoutsideofthisrange,pleaseobtainthermocouplesratedaccordingly.
Accuracyspecifiedformeteronly.Inputprobesaddadditionalmeasurementerror.GenerallyRTD
probesintroduceasmallererrorthantherm
ocouples.
TheabovespecificationsaretestedunderanenvironmentalRFFieldStrengthlowerthan3V/ Mand
afrequenc y lowerthan30MHz
SensorType Resolution Range Accuracy
TypeK
0.1
°
C
100.0to‐50.1
°
C
±(0.4% +1
°
C)
50.0to999.9
°
C
±
(0.4%+0.5°
C)
1°C
1000to1300
°
C
±(0.4% +1
°
C)
0.1
°
F
148.0to‐58.1
°
F
±(0.4%+1.8
°
F)
58.0to999.9
°
F
±(0.4% +1
°
F)
F
1000to2372
°
F
±(0.4% +2
°
F)
TypeJ
0.1
°
C
100.0to‐50.1
°
C
±
(0.4%+
C
)
50.0to999.9
°
C
±(0.4%+0.5
°
C)
C
1000to1150
°
C
±(0.4% +1
°
C)
0.1
°
F
148.0to‐58.1
°
F
±(0.4%+1.8
°
F)
58.0to999.9
°
F
±
(0.4%+
F
)
1
°
F
1000to2102
°
F
±(0.4% +2
°
F)
TypeT
0.1
°
C
100.0to‐50.1
°
C
±(0.4% +1
°
C)
50.0to400.0
°
C
±(0.4%+0.5
°
C)
0.1
°
F
148.0to‐58.1
°
F
±(0.4%+1.8
°
F)
58.0to752.0
°
F
±
(0.4%+
F
)
TypeE
0.1
°
C
100.0to‐50.1
°
C
±(0.4% +1
°
C)
50.0to900.0
°
C
±(0.4%+0.5
°
C)
0.1
°
F
148.0to‐58.1
°
F
±(0.4%+1.8
°
F)
58.0to999.9
°
F
±
(0.4%+
F
)
1
°
F
1000to1652
°
F
±(0.4% +2
°
F)
TypeR
1
°
C
0to600
°
C
±(0.5% +3
°
C)
601to1700
°
C
F
32to1112
°
F
±(0.5% +5
°
F)
1113to3092°F
TypeS
1
°
C
0to600
°
C
±(0.5% +3
°
C)
601to1500°C
F
32to1112
°
F
±(0.5% +5
°
F)
1113
to2732
°
F

Related product manuals