Operating Manual
5
1.2.2 Class1 Division 1
1.2.2.1 iCAM502 (units dated before 18
th
July 2013)
iCAM502 units with date codes before 18
th
July 2013 are not Class 1 Division 1 approved.
1.2.2.2 iCAM501U (all date codes) and iCAM502 (units dated after 18
th
July 2013)
(Taken from FM approvals certificate 3042108 amended July 12
th
2013)
Special conditions for use;
1. Must be used with Battery specified in the Operations Manual 331684
2. Temperature code is T4, T1=-20
o
C to 55
o
C with Duracell Plus or Procell MN1500
batteries, or T3C, Ta=-20
o
C to 50
o
C with Duracell Procell PC1500 or Energizer E91
batteries.
1.3 List of Notes
The notes supplied in this chapter provide information on the following.
Danger! / Warning!
o Possible hazard to life or health.
Caution
o Possible damage to property.
Important
o Possible damage to enclosure, device or associated equipment.
Information
o Notes on the optimum use of the device
1.3.1.1 Notes common to all iCAM501U and iCAM502 Units
Warning! Any connection to the iCAM50x USB port must NOT be made in a
hazardous area.
Warning! The user shall NOT mix different cells in ANY application
Warning! For ATEX/IECEx GROUP I applications ONLY, the changing of the cells
in the hazardous area or the transportation of spare cells into the
hazardous area is NOT PERMITTED.
Warning! For ATEX/IECEx GROUP II and Class 1 Division 1 applications ONLY, the
user shall not carry more than a maximum of 2 spare approved cells into
the hazardous area. Precautions shall be taken to prevent short
circuiting of un-protected cell terminals. Discharged cells shall be
removed from the hazardous area.
Warning! Warning! For ATEX/IECEx GROUP II applications ONLY the rated
ambient temperature range of the equipment is from -20
o
C to 60
o
C.