MANUAL THERMOCOAGULATION - EVRF
Date
Page 12 of 34 Revision N° : 18
The number of impulse per session.
Type BF device.
CE marking.
Read manual before use.
Allowed voltage.
Frequency.
Maximum power.
2.3 Output characteristics
This apparatus generates some high frequencies impulses whose characteristics are:
Frequency of the wave : 4 MHz.
Max voltage in output : 660V.
Max time of impulse : 0,8s.
Continuous mode.
2.4 Label explanation
1: Type BF apparatus.
2: Name + address of fabricant.
3: CE number.
4: Read operating manual before usage.
5: Model/apparatus name.
6: Serial number.
7: Technical specifications.
2.5 Functioning
The apparatus works in the following environment:
Room temperature between +10°C and + 40°C.
Relative dampness comprised between 30% and 80%, condensation comprised.
Atmospheric pressure comprised between 700 hPa and 1060 hPa.
The power supply cannot exceed a voltage of 230V~+-10% of the rated voltage.