Summary

Symbols and Warnings

Symbols Used in This Manual

Explains symbols used within the manual for warnings, cautions, and notes.

Symbols on Labeling and Packaging

Details symbols found on the product data plate and packaging for compliance and safety.

Intended Use and User Profile

Intended User Details

Defines required professional qualifications, skills, experience, and possible user handicaps.

Product Description and Variants

Device Operation Basics

Switching On;Off and Intensity Adjustment

Details how to power the device on/off and adjust light intensity using controls.

Color Temperature Adjustment

Describes how to adjust the color temperature of the light on tunable white versions.

Advanced Operation Features

Composave and Auto-On Settings

Explains Composave setting to avoid material polymerization and Auto-On for automatic lamp activation.

Theia Light and Remote Control

Covers switching the secondary Theia light and using the remote control.

Synchronized Room Light Operation

Details the procedure for creating a synchronized lighting system with Faro Room Light.

Preventive Maintenance and Checks

Routine Inspection Schedule

Outlines yearly checks for joint integrity, screw tightness, and component security.

Electrical and Light Performance Checks

Describes checks for oxidation, plate readability, enclosure integrity, electrical safety, and light output.

Cleaning and Sterilization Procedures

General Cleaning and Disinfection Warnings

Lists prohibited substances for cleaning to prevent wear, corrosion, and damage.

Specific Cleaning Instructions

Provides guidance on cleaning reflecting parabolas, head, and arms.

Handle Sterilization

Details handle removal, decontamination, disinfection, and sterilization parameters.

Troubleshooting and Technical Data

Error List and Indicators

Lists potential error codes, their descriptions, and associated acoustic indications.

Technical Specifications Overview

Provides detailed specifications including power, illuminance, color rendering, and radiation type.

Environmental Conditions

Storage and Transportation Conditions

Specifies environmental conditions for storing and transporting the device.

Operational Use Conditions

Defines the required environmental conditions for safe and effective device usage.

Overview

The EVA is a dental operating light designed for illuminating the oral cavity and structures of patients during dental procedures. It is intended for use in dental offices by dentists (all specializations), dental nurses, hygienists, and assistants. The device is positioned at a distance of 700mm from the operative area for optimal lighting.

Function Description:

The EVA dental light is available in two main product variants:

  • Sunlight Version: Features a light source at 5000 K with a "Sunlight" spectrum.
  • Tunable White Version: Offers a tunable white light source with selectable color temperatures of 4000 K, 5000 K, 5700 K, and a Composave setting at 2700 K, allowing the operator to choose the desired light color.

The device can be supplied with various configurations, including different mounting options, arm length combinations, Canbus Technology, remote cable for dental unit command integration, auto-on setting, Theia Technology (a secondary light source under the fixed arm), integrated 2K or 4K cameras, and RF for connection with FARO's room lights (Siderèa).

Important Technical Specifications:

  • Power Supply: 24 V ac ±10% 50/60Hz; 32 V dc ±10%.
  • Max Power (Dental Light Head): 24 V ac: 26 VA; 32 V dc: 14 VA.
  • Max Power (Dental Light with Theia Tech): 24 V ac: 40 VA; 32 V dc: 28 VA.
  • Recommended Fuses for Installation: 24 V ac: T1.6AL 250V; 32 V dc: T1.25AL 250V (for Dental Light Head). 24 V ac: T2AL 250V; 32 V dc: T1.25AL250V (for Dental Light with Theia Tech).
  • Protection against electrical hazard: Class II (Dental Light Head); Class I for Ceiling Mounting for Canada US Certification.
  • Classification against IEC 62471: Class 1 - Labelling Exempt.
  • Max Illuminance: > 45,000 lux.
  • Color Rendering Index (CRI): > 95.
  • Correlated Colour Temperature (CCT): Sunlight Version: 5000 K; Tunable White Versions: 4000 K - 5000 K - 5700 K; Composave setting: 2700 K.
  • Spot Dimension: 180 mm x 110 mm.
  • Max Hard Shadow ISO 9680: 10 mm x 5 mm.
  • Nature of the radiation: Non-ionizing Radiation.
  • Type of radiation: Visible light.
  • Storage and Transportation Environmental Conditions: Temperature from -20°C to +70°C, Relative humidity from 10% to 90%, Atmospheric pressure from 50 kPa to 106 kPa.
  • Use Environmental Conditions: Temperature from 10°C to 40°C, Max altitude: 2000 m, Relative humidity from 30% to 75%.
  • Overall Movements: Head rotations of 360°+90° and 110°.

Usage Features:

The EVA light offers an intuitive user interface with a joystick, sensor, and sterilizable handles.

  • Switching On/Off:
    • Joystick Version: Push right or left and release (1 beep).
    • Sensor Version: Bring hand towards the sensor up to 2 cm and move hand down (1 beep).
    • The device performs a self-diagnosis upon switching on, indicated by blinking blue, green, and red lights. It memorizes the last light intensity setting and color temperature (for Tunable White version).
  • Adjusting Light Intensity:
    • Joystick Version: Push left and keep pushed to increase intensity, push right and keep pushed to decrease intensity. Release when desired intensity is reached.
    • Sensor Version: Bring hand towards the sensor up to 2 cm and keep this distance to adjust intensity. Intensity changes continuously, decreasing until minimum and increasing until maximum.
    • Indicator light changes according to the illuminance level.
  • Changing Color Temperature (Tunable White Version):
    • Perform a double click on the joystick or sensor. Repeat until the desired color temperature is shown on the indicator light (2 beeps indicate Tk change).
    • Indicator strip colors: Green for 5000 K (Sunlight), Yellow for 4000 K (Tunable White), White for 5000 K (Tunable White), Blue for 5700 K (Tunable White).
  • Composave Setting (Tunable White Version):
    • This setting allows working with restorative material without polymerization.
    • Joystick Version: Press and release (option 1) or single click (option 2).
    • Sensor Version: Single click.
    • A wigwag beep is heard, and the indicator strip changes to orange. Exit by a single click from the Compo-Save Setting.
  • Minimum Intensity Setting (Solar Version):
    • Joystick Version: Press and release.
    • Sensor Version: Single click.
  • Auto-On Setting: When enabled, the lamps turn on automatically without a specific command when power is supplied.
  • Switching On/Off Theia (Secondary Light):
    • The secondary light on the fixed arm can be switched on/off and adjusted in synchro-mode with the operatory light using button (7) on the fixed arm.
    • If switched on after the operatory light, it synchronizes automatically.
    • If switched on with the operatory light off, it settles at maximum intensity.
    • Theia light cannot be adjusted independently in synchro mode.
  • Remote Control: The dental light can be controlled from the dental unit's control panel (refer to the dental unit's instructions).
  • Syncro Mode with FARO Room Light:
    • The device can be paired with FARO Room Lights via Radio Frequency (RF) to create a synchronized lighting system.
    • Pairing Procedure: Switch on Room Light (starts searching for 60s). Within 60s, press "Pairing" button on dental lamp for 4-6s. Room Light's blue LED activates. Confirm pairing by pressing the programming button on the Room Light's remote control within 60s.
    • De-activation: Press Synchro button for 2-4s, then release (sound signal, blue LED on Room Light goes out).
    • When synchronized, the Room Light's blue LED is steady on. The remote control remains enabled for manual illumination changes. If the Dental Light is off, the Room Light remains switched on.

Maintenance Features:

Preventive maintenance and routine checks should only be performed by a Service Engineer according to the Manufacturer's Service Manual.

  • Checks (Yearly):
    • No plays or space gaps between junction points (1, 2, 3, 4).
    • Tightness and integrity of screws (5, 6) and nuts under carters (1, 2, 3).
    • Visual inspection for absence of oxidation in joints, arms, or plastic parts.
    • Readability of the main plate.
    • Damages on enclosure and plastic joints integrity.
  • Electrical Safety (Every two years): According to EN 62353 (Dielectric strength, Current Leakage).
  • Light Checks (Every two years): With a spectroradiometer, check Max Luminance (>35000 lux), CRI (>85), and Radial power on blue light (<100 W/m2).

Cleaning and Disinfection:

  • General Cleaning: Use a soft cloth (cotton or absorbent cotton) with ethyl alcohol or water-alcohol based disinfectants (70% isopropyl alcohol or ethanol).
  • Forbidden Substances: Abrasive, corrosive, acids, substances containing chlorine or chloride ions, phosphorous or phosphorous ions, detergents with Trilene base, petrol, white spirit, or similar. For plastic parts, avoid ammonium hydroxide, sodium hydroxide, hydrogen peroxide, ammonium chloride, methylene chloride, methyl alcohol. Do not spray chemicals directly on the device or use wet wipes without rinsing.
  • Recommended Disinfectants: Durr FD366 Sensitive, Faro Perflex Advanced.
  • Reflecting Parabolas: Clean with a soft cloth in cotton or absorbent cotton with ethyl alcohol. Wear gloves to avoid fingerprints. Do not spray detergent directly. Avoid detergents with surfactants or water-repellents that leave streaks.
  • Head: Clean with a soft cloth in cotton wetted with disinfectant solution, squeezing excess liquid.
  • Arms: Use a cloth soaked in approved disinfectant, squeezing excess liquid.

Sterilization of Handles:

  • Warning: Handles are not supplied sterile and must be sterilized before each patient.
  • Removal: Unscrew knob "A" and remove from support.
  • Decontamination and Disinfection: Before sterilization, decontaminate and disinfect handles. Recommended products: Faro Perflex Advanced, Durr FD366 Sensitive.
  • Warning: Handles cannot be disinfected by thermo-disinfection (danger of plastic breaking).
  • Sterilization: Handles must be packaged according to EN 868-5. They can be sterilized with standard cycles 121°C/134°C up to 200 cycles or until loss of mechanical performance.
    • Cycle EN 13060 B: Temperature 121°C, Pressure 207 kPa, Holding Time Minimum 15 min.
    • Cycle EN 13060 B: Temperature 134°C, Pressure 308 kPa, Holding Time Minimum 3 min.

The device is a Medical Device according to Regulation (EU) 2017/745 and is double insulated (Class 2 device against electrical risk). It carries the CE mark for Conformitè Europe and is WEEE compliant. The product has a Service Life of 10 Years and comes with a 24-month warranty from installation (max 30 months from manufacturing date).