5 PREVENTIVE MAINTENANCE AND ROUTINE CHECKS
Only Service Engineer are allowed to perform corrective Maintenance and replacement of any part of the device, according to Manufacturer’s Service
Manual.
No plays or space gaps between the
junction points (points 1, 2, 3, 4)
Screws of connection points must
be tightened and integer:
screws 5
Screw 6.
The nuts of connection points
under carter 1, 2 must be well
secured and safety screws intact.
The screws under carter 3 must be
well secured.
Carter 3 Carter 2 Carter 1
Check the absence
of any oxidation into joints, arms or
plastic parts.
Check the main plate can be read
Check of damages on enclosure and
plastic joints integrity.
Electrical Safety according EN
62353
1. Dielectric strength
2. Current Leakage
Use the parameters defined into IEC 60601-1
With a spectroradiometer check the values for:
Max Luminance: >35000 lux
CRI > 85
Radial power on blue light: <100 W/m2