FT-70/75 Fill Level Inspectors Operator’s Guide
2-18
4. Find your type of Missing Lid Sensor In the table below and
adjust the upper jam nut to raise or lower the sensor to the
proper height.
5. When you achieve the proper height, tighten the lower jam nut
to secure the sensor positioning.
6. If you will not be performing further alignment procedures,
reactivate the X-ray Generator. For more information see “The
Safety Interlock Switch” on page 2-4.
Description (Lid Type) Maximum Height Above Lid
Nonferrous (Aluminum) 0.4 inch (10 mm)
Photo-Optic Sensor (All) 0.7 inch (18 mm)
Extended Range (Aluminum)
Extended Range (Steel)
0.5 inch (13 mm)
0.88 inch (22 mm)
Standard 0.25-0.38 inch (6-9 mm)
Industrial Dynamics Company, Ltd. Document 40274-0600