EasyManua.ls Logo

Fisher Scientific accumet XL150 accumet XL200 accumet XL250 accumet XL500 accumet XL600 User Manual

Fisher Scientific accumet XL150
accumet XL200
accumet XL250
accumet XL500
accumet XL600
63 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
1
Reference Guide
Fisher Scientific
accumet
®
Excel (XL)
Benchtop Meters
XL150 XL200 XL250 XL500 XL600
68X623601 Rev 0 October 2012
Question and Answer IconNeed help?

Do you have a question about the Fisher Scientific accumet XL150 accumet XL200 accumet XL250 accumet XL500 accumet XL600 and is the answer not in the manual?

Fisher Scientific accumet XL150 accumet XL200 accumet XL250 accumet XL500 accumet XL600 Specifications

General IconGeneral
mV Resolution0.1 mV
mV Accuracy±0.2 mV
mV Range±2000 mV
Temperature Range-5.0 to 105.0°C
Temperature Resolution0.1°C
Conductivity Resolution0.01 μS/cm to 0.1 mS/cm
Salinity Range0 to 80 ppt
Resistivity Range0 to 100 MΩ·cm

Summary

1. INTRODUCTION

Intended Use

Details the intended use of the accumet XL series meters for water-based electrochemical parameters in laboratory environments.

2. TOUCH SCREEN & ICONS

Touch Screen Operation and Maintenance

Explains touch screen usage, preventing damage, and routine cleaning procedures.

Administration and System Setup Icons

Covers icons for user management (Admin) and instrument configuration (System Setup).

Meter Operation Icons

Details icons for instrument status (About), general controls (Mode, Setup), and user management.

Data Management Icons

Covers icons for measurement modes, printing, data logging, and viewing stored data.

Auxiliary Controls and Buttons

Includes icons for stirrers, help, and buttons for standardization actions like Confirm and Cancel.

3. CONNECTIONS

Rear Panel Port Overview

Illustrates and labels all input/output ports on the rear of the XL600 meter.

Side Panel Ports

Details USB, RS-232, TEMP, CELL, POWER, BNC, GROUND, STIRRER B, and RJ45 ports.

BNC Electrode Connection Note

Instructions for removing and using the BNC shorting cap for pH, ORP, or Ion electrodes.

4. ELECTRODE BRACKET & ARM ASSEMBLY

Bracket Attachment Procedure

Instructions for connecting the metal electrode bracket to the meter base using a Phillips screwdriver.

Wall Mounting Consideration

Notes that the electrode arm should be removed from the bracket if the instrument is wall mounted.

Arm and Bracket Assembly

Details attaching the electrode arm to the bracket and tightening the wing nut.

5. POWER ON/OFF

Powering On the Unit

Procedure to power on the meter by pressing the on/off key for three seconds.

Powering Off the Unit

Procedure to power off the meter by pressing the on/off key for three seconds and confirming shutdown.

6. SYSTEM SETUP

Accessing System Setup Menu

How to access System Setup by selecting MODE and then the 'Sys. Setup' icon.

Customizable System Settings

Lists settings that can be customized, including Stirrer Speed, Brightness, Printer Option, and language.

Software Update Information

Advises checking www.fishersci.com/accumet for free instrument software updates.

7. ADMINISTRATION & USER ID

Benefits of User IDs

Explains advantages of User IDs like security, separation of data, and detailed data logging.

User Login Process

Describes the automatic User Login dialog box that appears upon instrument startup.

User Profile Management

Step-by-step guide to access Administration, set up user profiles, and add new User IDs.

Creating and Managing User Profiles

Details on entering User ID, Password, Company Name, and confirmation for adding new users.

User Profile Screen Function Keys

Explains the functions of Add, Delete, Modify, Exit, Apply To Current/All User, Default Setting, Delete User Data.

8. PARAMETER SETUP: SAMPLE ID

Sample ID Usage for Data Logging

Explains that a Sample ID is required for logging data or activating timed printing.

Sample ID Assignment Options

Describes Manual, Sequence, and None options for Sample ID assignment.

9. PARAMETER SETUP: AUTO READ MODE

Auto Read Mode Settings

Details Auto and Manual settings for the Auto Read option, affecting measurement freezing.

Auto Read Inactivity for ISE Methods

Informs that Auto Read is inactive for certain ISE methods like Known Addition.

10. PARAMETER SETUP: STABILITY CRITERIA / STABLE INDICATOR

Stability Criteria Options

Lists options for pH, mV, ISE, and Conductivity/Resistivity/TDS/Salinity stability criteria.

Adjusting Stability Indicator Speed

Explains how to adjust the STABLE indicator appearance using FAST, MEDIUM, or SLOW settings.

Tip: pH Resolution for Stabilization

Suggests selecting lower pH resolution to decrease stabilization time.

11. PARAMETER SETUP: DEFAULT TEMPERATURE

Purpose of Default Temperature Setting

Explains how default temperature settings eliminate errors and improve pH accuracy.

Setting the Default Temperature Value

Step-by-step guide to set default temperature units and value using the keypad.

ATC Probe Override Note

Informs that an ATC probe will override any manually entered default temperature value.

12. PARAMETER SETUP: ISOPOTENTIAL POINT

Understanding Isopotential Point

Defines isopotential point and its importance for pH and Ion Selective Electrodes.

Setting the Isopotential Point

Instructions to set the isopotential point in mV using the numeric keypad.

Applying Temperature Compensation for Ions

Explains how to apply ATC by entering the Isopotential Point for ISE experiments.

13. PARAMETER SETUP: ALARM LIMITS

Alarm Limit Functionality

How alarm limits alert users to High or Low value excursions with visual and audible signals.

Setting High and Low Alarm Limits

Steps to enable/disable alarms and set Low and High limit values using the keypad.

Tip: Configuring Single Alarm Limits

Advises setting only High or Low alarm by adjusting the respective limit value.

14. PARAMETER SETUP: PRINT CRITERIA

Print Criteria Configuration

Allows selection of criteria to be printed along with measurement data.

Setting Print Interval Options

Steps to edit print criteria, select ON/OFF for details, and choose Log Data or Timed interval.

15. PARAMETER SETUP: DATA STORAGE CRITERIA

Data Storage Configuration

Enables selection of criteria to include when saving data into memory.

Setting Data Logging Interval

Steps to edit data storage, select ON/OFF for details, and choose MANUAL, STABLE, or TIMED logging interval.

16. PARAMETER SETUP: DISPLAY CRITERIA

Display Criteria Configuration

Allows selection of criteria to appear during measurement.

Setting Display Resolution

Steps to edit display criteria, select ON/OFF for details, and set display resolution.

17. VIEWING & EXPORTING STORED DATA (LOG VIEW)

Accessing and Viewing Stored Data

Allows examination of specific data points stored in the meter, up to 2000 points per user ID.

Exporting Data to USB Drive

Instructions to connect a USB drive and select 'Export' to save data.

18. GRAPHING

Real-Time Data Graphing

View real-time data changes on a graph, plotting time in seconds.

Graph Interaction Controls

Details on Show Graph, Start/Stop Plotting, Zoom In/Out, and Hide Graph functions.

19. STANDARDIZATION

Purpose and Importance of Standardization

Explains maximizing accuracy by storing values of known standards for sample measurement.

Step-by-Step Standardization Guide

Step-by-step guide for performing a typical standardization using known standard solutions.

Standardization Control Buttons

Describes the Confirm, Clear, and Cancel buttons used during the standardization process.

20. TEMPERATURE OPERATION

Temperature Sensor Calibration

Recommends temperature calibration upon electrode replacement or when readings are suspect.

Standardizing Temperature Readings

Procedure to standardize the temperature by selecting 'Temp Std.' and entering a new value.

21. PH OPERATION

pH Buffer Group Selection

Lists available buffer groups (USA, DIN, NIST, etc.) for pH standardization.

Custom pH Buffer Setup

Instructions for entering custom pH buffer values for standardization.

pH Buffer Recognition Modes

Explains Automatic and Manual buffer recognition modes for standardization.

pH Standardization Best Practices

Provides tips on best practices for pH standardization, buffer reuse, and electrode care.

22. MILLIVOLT OPERATION

Millivolt Measurement Modes

Describes measuring oxidation/reduction potential (ORP/Redox) in absolute or relative mV mode.

Relative mV Standardization Procedure

Procedure to adjust the displayed value in relative mV mode using SET or DEFAULT.

23. ION: SETUP & OPERATION

Ion Method Selection Guide

Guide to choosing the desired Ion Method from Direct Reading, Known Addition, and others.

Ion Measurement Method Descriptions

Details Direct Reading, Blank Offset, Known Addition, Known Subtraction, Analate Addition, Analate Subtraction methods.

Electrode Type Configuration

Instructions for choosing the electrode type and its impact on calculations and standardization.

Ion Measurement Units Selection

How to select the units for reporting ion concentration, such as ppm or mg/L.

Ion Temperature Compensation Setting

Option to set ATC to YES or NO for ion concentration measurements.

Setting Isopotential Point for Ion Measurement

Steps to set the isopotential point for Ion Selective Electrodes in mV.

24. ION: DIRECT READING METHODS

Direct Reading with Standards Method

Procedure for measuring ion concentration using known standards.

Direct Reading with Blank Offset Method

Method for measuring low ion concentrations by subtracting blank values.

Tip: STABLE Indicator for Direct Reading

Notes that STABLE appears after two-point standardization; dashes indicate insufficient standardization.

Standard or Blank Reading Prompt Response

Explains how to respond to the prompt asking if the reading is STANDARD or BLANK.

25. ION: INCREMENTAL METHODS

Usefulness of Incremental Methods

Highlights usefulness for samples with complex ionic backgrounds or varying temperatures.

Standardization for Incremental Methods

Requirement to standardize with at least two standards to establish a slope value.

26. ION: KNOWN ADDITION METHOD

Known Addition Procedure Steps

Step-by-step guide for performing the Known Addition method, including parameter entry.

Known Addition Result Display Details

Details the information displayed after a Known Addition calculation, like slope and readings.

27. ION: KNOWN SUBTRACTION METHOD

Known Subtraction Procedure Steps

Step-by-step guide for performing the Known Subtraction method, similar to Known Addition.

Known Subtraction Result Display Details

Details the information displayed after a Known Subtraction calculation.

28. ION: ANALATE ADDITION METHOD

Analate Addition Procedure Steps

Step-by-step guide for performing the Analte Addition method.

Analate Addition Result Display Details

Details the information displayed after an Analate Addition calculation.

29. ION: ANALATE SUBTRACTION METHOD

Analate Subtraction Procedure Steps

Step-by-step guide for performing the Analate Subtraction method.

Analate Subtraction Result Display Details

Details the information displayed after an Analate Subtraction calculation.

30. CONDUCTIVITY / RESISTIVITY OPERATION

Conductivity Standard Recognition Options

Explains Auto, Manual, Single Point, and Multi Point standard recognition for conductivity.

Conductivity and Resistivity Ranges

Lists conductivity and resistivity values for Ranges 1 through 4.

Cell Constant Selection Guide

Guidance on selecting the nominal cell constant for conductivity electrodes.

Default Temperature for Conductivity

States the default temperature for conductivity measurements (25 °C / 77 °F).

Temperature Coefficient Adjustment

Explains adjusting the temperature coefficient for conductivity measurements.

Reference Temperature Setting

How to set the reference temperature for automatic temperature compensation.

31. TDS OPERATION

TDS Measurement Principle Explained

Explains TDS determination by multiplying electrical conductivity by a TDS conversion factor.

TDS Conversion Factors Table

Provides TDS values and conversion factors for common salts.

TDS Factor Adjustment Procedure

How to adjust the TDS factor in the TDS Setup menu.

TDS Measurement Ranges

Lists TDS values for Ranges 1 through 4, calculated using the TDS factor.

32. SALINITY OPERATION

Salinity Measurement Principle Explained

Describes salinity measurement using a non-linear practical salinity curve based on NaCl.

Salinity Measurement Ranges

Lists salinity values for Ranges 1 through 4.

33. DISSOLVED OXYGEN OPERATION

Dissolved Oxygen Measurement Modes

Lists modes measured: DO, BOD, OUR, SOUR. Each mode is independent.

DO Calibration Mode Options

Details % Saturation, mg/L, and mbar calibration modes.

Calibrating the Internal Barometer

Procedure to calibrate the internal barometer for accurate barometric pressure readings.

Setting Salinity for Dissolved Oxygen

How to set salinity to compensate for its impact on dissolved oxygen concentration.

Tip: DO Probe Warm-up Requirement

Advises allowing the DO probe to warm up for at least 10 minutes before use.

One Point Auto Standardization - % Saturation

Procedure for one-point auto standardization assuming 100% saturation.

Two Point Auto Standardization - % Saturation

Procedure for two-point auto standardization at 0% and 100% saturation.

One Point Auto Standardization - mg/L

Procedure for one-point auto standardization for mg/L.

Two Point Auto Standardization - mg/L

Procedure for two-point auto standardization at 0 mg/L and 8.26 mg/L.

One Point Manual Standardization - mg/L or % Saturation

Procedure for one-point manual standardization for mg/L or % saturation.

34. BOD OPERATION

BOD Configuration Settings Explained

Details settings for BOD seed minimum Delta DO, seed endpoint, sample Delta DO, and sample endpoint.

35. OUR OPERATION

OUR Configuration Settings Explained

Details OUR dilution factor, minimum/maximum time, and minimum starting/ending DO.

36. SOUR OPERATION

SOUR Configuration Settings Explained

Details SOUR dilution factor, minimum/maximum time, starting/ending DO, and solids weight.

37. FIRMWARE UPGRADE

Determining Current Software Version

Methods to check the current software version of the instrument.

Checking for Available Software Updates

Instructions to check www.fishersci.com/accumet or email for software updates.

Saving Software Update to USB Drive

Steps to save new software to the root folder of a USB drive.

Initiating the Firmware Upgrade Process

Procedure to connect the USB drive and initiate the software upgrade process.

Monitoring Software Update Progress

Information window indicates 'Please wait' during the software update.

Completing Firmware Installation

Steps to confirm installation completion and shut down the instrument.

38. SPECIFICATIONS

pH Specifications Details

Details range, resolution, accuracy, buffer sets, and inputs for pH measurements.

mV Specifications Details

Details range, resolution, accuracy, and offset adjustment for mV measurements.

Ion Specifications Details

Details range, resolution, units, accuracy, and calibration points for Ion measurements.

Temperature Specifications Details

Details range, resolution, accuracy, and calibration for temperature.

Conductivity Specifications Details

Details range, resolution, accuracy, calibration points, cell constant, and normalization.

TDS and Salinity Specifications Details

Details range, resolution, accuracy, and calibration points for TDS and Salinity.

Dissolved Oxygen Specifications Details

Details range, resolution, relative accuracy, and barometric pressure for DO.

Other Instrument Specifications

Details display, output, memory, alarms, meter size, weight, and power.

Instrument Operating Conditions

Specifies operating ambient temperature, humidity, storage conditions, pollution, and enclosure rating.

Power Adapter Operating Conditions

Specifies operating ambient temperature, humidity, and storage conditions for the power adapter.

39. REPLACEMENTS & ACCESSORIES

XL Series Meter Kits and Models

Lists available meter kits for XL150, XL250, XL200, XL500, XL600 models with included components.

Electrodes, Probes, and Cables

Lists various pH, ATC, conductivity, and DO electrodes and probes with catalog numbers.

Meter Accessories

Lists accessories like stirring probes, mice, and replacement paddles with catalog numbers.

Conductivity Cells and Accessories

Lists different conductivity cells (2-cell, 4-cell) with varying cell constants and body materials.

Calibration Kits and Printers

Lists conductivity calibration kits, electrode support arms, printers, and power supplies.

40. WARRANTY

Warranty Terms and Conditions

Details warranty period for meters and electrodes, repair or replacement policy.

Disclaimer of Implied Warranties

Expressly disclaims other warranties like merchantability and fitness for a particular purpose.

41. RETURN OF ITEMS

Return Goods Authorization Process

Explains the requirement to obtain an RGA from Customer Service before returning items.

Packing, Insurance, and Restocking Fees

Advises careful packing, insurance, and notes that restocking charges apply to unauthorized returns.

42. NOTICE OF COMPLIANCE

FCC and Radio Interference Notice

States compliance with FCC rules for Class A computing devices and potential interference.

WEEE Compliance Information

Information on compliance with the European Union's Waste Electrical & Electronic Equipment Directive.

43. DECLARATION OF CONFORMITY

Product Conformance to Standards

Declares that specific accumet XL products conform to EMC and Safety standards.

Related product manuals