14. EMC-Declaration of conformity
The device has been tested and homologated in accordance with EN 60601-1-2 for EMC. This
does not guarantee in any way that this device will not be affected by electromagnetic
interference Avoid using the device in high electromagnetic environment.
14.1 Technical Description Concerning Electromagnetic Emission
Table 1: Guidance & declaration - electromagnetic emissions
The model Rooter
®
X3000 is intended for use in the electromagnetic environment specified below. The customer
or the user of the model Rooter
®
X3000 should assure that it is used in such an environment.
Electromagnetic environment - guidance
CISPR 11
The model Rooter
®
X3000 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
CISPR11
The model Rooter
®
X3000 is suitable for used in all
establishments, including domestic establishments and those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
lEC 61000-3-2
/ flicker emissions
lEC 61000-3-3
14.2 Technical Description Concerning Electromagnetic Immunity
Table 2: Guidance & Declaration - electromagnetic immunity
The model Rooter
®
X3000 is intended for use in the electromagnetic environment specified below. The customer
or the user of the model Rooter
®
X3000 should assure that It is used in such an environment.
test level
Electromagnetic environment -
guidance
discharge (ESD)
lEC 61000-4-2
±2, ±4, ±8, ±15kV air
±2, ±4, ±8, ±15kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
transient/burst
IEC 61000-4-4
supply lines
output lines
lines
Mains power quality should be that of
a typical commercial or hospital
environment.
lEC 61000-4-5
±0.5, ±1, ±2kV line to
earth
±0.5, ±1, ±2kV line to
earth
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
lines
(>95% dip in UT.)
for 0.5 cycle
(>95% dip in UT.)
for 0.5 cycle
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the models
Rooter
®
X3000 requires continued
operation during power mains