FLIR CM275 USER MANUAL Document Identifier: CM275-en-US_AC
37
13. Specifications
13.1 General specifications
Display counts: 0~6000
Measuring rate: 3 times per second
Over-range indication: OL or OL
Auto power off: Programmable: OFF, 1, 2, 5, or 10 (default) minutes
Low battery indicators: and low battery informaton screens appear
Power requirement: 3 × 1.5 V AA or TA04 rechargeable battery
Approximate battery life for thermal imager:
hours: Energizer L91 Lithium (Li/FeS
2
hours: Optional Rechargeable Battery: Li-Polymer; FLIR PN: TA04-KIT
Calibration: 1-year calibration cycle
Operating temperature:
32 to 86°F (0 to 30°C) ( 80% RH)
86 to 104°F (30 to 40°C) ( 75% RH)
104 to 122°F (40 to 50°C) ( 45%RH)
Storage temperature: 4 to 140°F (-20 to 60°C)
Relative Humidity: 0~80% RH (batteries not installed)
Temperature coefficient: 0.2 × (specified accuracy)/°C, <64.4°F (18°C), >82.4°F (28°C)
Operating altitude: (2000m)
Jaw opening: 1.38in (35mm)
Pollution degree: 2
IP Rating IP40
Dimensions: (D × W × L): 1.91 × 3.82 × 10.04 in (48.5 × 97 × 255 mm)
Weight: 16.2 oz (460g) without batteries
Agency Approvals: UL, CE, RCM
Over-voltage category: EN 61010-1 CAT IV-600 V, CAT III-1000 V, EN 61010-2-032
CAT Application field
III Distribution circuits, machinery, main switching devices close to switchgears,
industrial installations and high current close to distribution circuits
IV Installation sources, utility transformers, all outside conductors, counters, protective
devices on primary sides and electricity meters