Technical data13
Ethernet
Ethernet, image streaming 16-bit 320 × 240 pixels
Signal linear
Temperature linear
Radiometric
Ethernet, power Power over Ethernet, PoE IEEE 802.3af class 0.
NOTE: In cameras manufactured before 2013,
due to an error in the implementation of power
over Ethernet, in some rare cases the camera will
not be powered. In such cases, power the camera
using the external power cable, or modify the
camera according to Service bulletin SB14-006.
For modification, please contact your local service
department. See http://support.flir.com/service for
contact details.
Ethernet, protocols TCP, UDP, SNTP, RTSP, RTP, HTTP, ICMP, IGMP,
ftp, SMTP, SMB (CIFS), DHCP, MDNS (Bonjour),
uPnP
Digital input/output
Digital input, purpose
Image tag (start/stop/general), Input ext. device
(programmatically read)
Digital input
2 opto-isolated, 10–30 VDC
Digital output, purpose As function of ALARM, Output to ext. device (pro-
grammatically set)
Digital output 2 opto-isolated, 10–30 VDC, max 100 mA
Digital I/O, isolation voltage 500 VRMS
Digital I/O, supply voltage 12/24 VDC, max 200 mA
Digital I/O, connector type
6-pole jackable screw terminal
Composite video
Video out
Composite video output, PAL and NTSC
compatible
Video, standard
CVBS (ITU-R-BT.470 PAL/SMPTE 170M NTSC)
Video, connector type
Standard BNC connector
Power system
External power operation 12/24 VDC, 24 W absolute max
External power, connector type 2-pole jackable screw terminal
Voltage Allowed range 10–30 VDC
Environmental data
Operating temperature range –15°C to +50°C (+5°F to +122°F)
Storage temperature range –40°C to +70°C (–40°F to +158°F)
Humidity (operating and storage) IEC 60068-2-30/24 h 95% relative humidity +25°C
to +40°C (+77°F to +104°F)
EMC
EN 61000-6-2:2001 (Immunity)
EN 61000-6-3:2001 (Emission)
FCC 47 CFR Part 15 Class B (Emission)
Encapsulation IP 40 (IEC 60529)
Shock 25 g (IEC 60068-2-27)
Vibration 2 g (IEC 60068-2-6)
#T559498; r.22370/22370; en-US