412355A-EU V3.1 4/08
6
Specifications
General Specifications
Display 9999 count LCD
Meter Power 9 volt battery or 9V AC adaptor
Auto Power OFF Meter automatically powers off after 10 minutes of
inactivity
Current output capability 24mA at 1000 ohms
Operating Temperature 5ºC to 40ºC (41ºF to 104ºF)
Storage Temperature -20
o
C to 60
o
C (-4
o
F to 140
o
F)
Operating Humidity Max 80% up to 31ºC (87ºF) decreasing linearly to 50% at
40ºC (104ºF)
Storage Humidity <80%
Operating Altitude 7000ft. (2000meters) maximum
Dimensions 3.8 x 4.7 x 1.8” (96 x 118 x 45mm) folded
Weight 12 oz. (340g)
Accessories Supplied 9V battery, AC adaptor and calibration cable with spade
lugs
Range Specifications
Mode Function Range Resolution Accuracy (% of reading)
0 to 2000mV 1mV Voltage
2.00 to 20.00V 0.01V
Current 0.00 to 50.00mA 0.01mA
Measure
% -25.0 to 230.0% 0.1%
0 to 2000mV 1mV Voltage
2.00 to 20.00V 0.01V
Current 0.00 to 24.00mA 0.01mA
Source
% -25.0 to 125.0% 0.1%
± (0.075% + 1 digit) or ± 3
digits, whichever is greater
Copyright © 2008 Extech Instruments Corporation (a FLIR company)
All rights reserved including the right of reproduction in whole or in part in any form.