DCO65U-E SC2500 User Manual Page 29
video monitor remain black?
No The problem is solved.
YesContact FLIR Systems After Sales
Service
No Live image in Altair
Is the camera correctly connected and powered?
No remake connections
Yes Does Cirrus communicate with the
camera?
No Contact FLIR Systems After Sales
Service
Yes Check that external trigger is
deactivated. Is the problem solved?
No Contact FLIR Systems After Sales
Service
No image in external
triggering mode
Is the sync signal that is connected to the
camera LVTTL (0 3.3V) under 50 Ohms ?
No Adjust the input signal
Yes Is the frequency of the signal in the
range of the camera (from 3Hz to the MAX
frequency proposed by CIRRUS/ALTAIR)
No Adjust the frequency of the signal
YesReplace the external signal by a signal
given by a function generator setup to
square signal from 0 to 3.3V under
50Ohms and 25Hz. Do you get an
image ?
NoContact FLIR Systems After Sales
Service
YesCheck more precisely your
synchronization signal. Eg. : 50 HZ
noise, low voltage under 50 Ohms,
Rising time to long, etc.