EasyManua.ls Logo

FLIR TG275 - Measurement Specifications; Measurement Analysis Specifications; Configuration Specifications

FLIR TG275
38 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
10
10.4 Measurement Specifications
Object temperature range -13 ~ 1022 (-25 ~ 550)
Accuracy at ambient temperature: 59 ~
95 (15 ~ 35)
-13 ~ 32 (-25 to 0): ± 7.0
3.0)
32 ~ 122 (0 ~ 50): ± 5 (±2.5)
122 ~ 212 (50 ~ 100): ± 3
1.5)
213 ~ 932 (100 ~ 500): ± 2.5%
932 ~ 1022 (500 ~ 550): ± 3.0%
IR Temperature resolution 0.2 (0.1)
Reading repeatability ± 1 % of reading
Response time 150 ms
IR thermometer measurement Continuous scanning
Minimum measurement distance 0.85 ft. (0.26 m)
10.5 Measurement Analysis Specifications
Spot meter Center spot (cross-hairs); Programmable
ON/OFF
Color display palettes Iron, Rainbow, White-hot, Black-hot, Arctic,
and Lava
10.6 Configuration Specifications
Set-up commands Local adaptation of units, language, date
and time formats
Emissivity adjustment 3 presets plus a custom adjustment utility
(0.1 ~ 0.99)
Languages Czech, Danish, Dutch, English, Finnish,
French, German, Greek, Hungarian, Italian,
Japanese, Korean, Norwegian, Polish, Por-
tuguese, Russian, simplified Chinese,
Spanish, Swedish, traditional Chinese,
Turkish
Firmware updates User manageable (instructions included in
this user manual)
#NAS100009; r. AJ/63868/63869; en-US