1652C/1653B/1654B
Calibration Manual
6
Continuity (R
LO
)
Limit Value
Maximum Display
Value
Limit Value
Maximum Display
Value
0.2 0.16 3 2.68
0.3 0.25 4 3.58
0.4 0.34 5 4.48
0.5 0.43 6 5.38
0.6 0.52 7 6.28
0.7 0.61 8 7.18
0.8 0.7 9 8.08
0.9 0.79 10 8.98
1 0.88 20 17.98
2 1.78 30 26.8
Loop Tests (Z
I
)
Loop Z
I
Hi Current
Loop Z
I
No Trip
Loop Z
I
Loop R
E
Limit
Value
Maximum
Display
Value
Limit
Value
Maximum
Display
Value
Limit
Value
Maximum
Display
Value
Limit
Value
Maximum
Display
Value
0.20 0.14 - - 3 2.53 3 2.72
0.30 0.23 - - 4 3.38 4 3.62
0.40 0.32 0.40 0.28 5 4.23 5 4.52
0.50 0.41 0.50 0.37 6 5.08 6 5.42
0.60 0.50 0.60 0.45 7 5.93 7 6.32
0.70 0.59 0.70 0.54 8 6.78 8 7.22
0.80 0.68 0.80 0.62 9 7.63 9 8.12
0.90 0.77 0.90 0.71 10 8.48 10 9.02
1.00 0.86 1.00 0.79 20 16.98 20 18.02
1.10 0.95 1.10 0.88 30 25.3 30 27.2
1.20 1.04 1.20 0.96 40 33.8 40 36.2
1.30 1.13 1.30 1.05 50 42.3 50 45.2
1.40 1.22 1.40 1.13 60 50.8 60 54.2
1.50 1.31 1.50 1.22 70 59.3 70 63.2
1.60 1.40 1.60 1.30 80 67.8 80 72.2
1.70 1.49 1.70 1.39 90 76.3 90 81.2
1.80 1.58 1.80 1.47 100 84.8 100 90.2
1.90 1.67 1.90 1.56 200 169.8 200 180.2
2.00 1.76 2.00 1.64 300 253 300 272
- - - - 400 338 400 362
- - - - 500 423 500 452
- - - - 600 508 600 542
- - - - 700 593 700 632
- - - - 800 678 800 722
- - - - 900 763 900 812
- - - - 1000 848 1000 902