Electrical Installation Tester
Operation
19
Table 9. Error Codes (cont.)
Error Condition Code Solution
Fault Voltage 4
Check the voltage between N and PE. RCD, socket
test, U
L
is exceeded. Loop test no trip >10 V.
Excessive Noise 5
Turn off all appliances (Loop, RCD measurements) or
move the earth stakes (earth measurement).
Excessive Probe
Resistance
6
Put the stakes deeper into the soil. Tamp down the
soil directly around the stakes. Pour water around the
stakes but not at the earth ground under test.
Data Memory 9
The data memory is inconsistent. Download and save
all data to a PC and clear all memory in the Tester. If
the error persists, return the Tester to a Fluke Service
Center.