EasyManua.ls Logo

Fluke 1663 - Page 30

Fluke 1663
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrical Installation Tester
Operation
21
Table 10. Power-On Options (cont.)
Push
Button
Power-On
Option
Description
Fault voltage
limit
Toggles the fault voltage between 25 V and 50 V. The
default setting is 50 V.
Serial
Number
Primary display shows the initial four digits and the
secondary display shows the next three digits.
A
Continuity
beeper
Turn on and turn off the beeper. The default setting is
bEEP on.
Auto Start
Automatic test start toggle. Simultaneously press and
the UP cursor. When turned on, the unit starts an RCD or
loop test if mains voltage is detected. You do not need to
press . The default setting is AUSt oFF.
0 Hz/128 Hz
No Trip Loop test measurement frequency toggle.
Simultaneously press and the DOWN cursor. Use
0 Hz if the RCD under test has high impedance with the
higher frequency. The default setting is 128 Hz.
Note
0 Hz is not available in the Auto Test
Sequence.

Table of Contents

Related product manuals