4 Fluke Corporation Fluke 345 Power Quality Clamp Meter
Display
Color transmissive LCD 320 x 240 pixels (70 mm diagonal) with 2 level backlight
Power supply
Battery type 1.5 V Alkaline AA MN 1500 or IEC LR6 x 6
Battery life typically > 10 hours (backlight on full)
> 12 hours (backlight reduced)
Battery Eliminator BE345
Input 110 V/230 V, 50/60 Hz
Output 15 V dc, 300 mA
Ambient conditions (For indoor use only)
Reference conditions All accuracies stated at 23 °C ± 1 °C (73.4 °F ± 1.8 °F)
Operating temperature 0 °C to 50 °C (32 °F to 122 °F)
Temperature coefficient of current
± 0.15 % of rdg per °C
Temperature coefficient of voltage
± 0.15 % of rdg per °C
Maximum relative humidity 80 % for temperatures up to 31 °C (87 °F) decreasing linearly to 50 % relative humidity at
40 °C (104 °F)
Maximum operating altitude 2000 m
Electrical safety
Safety IEC 61010-1 600 V CAT IV, double or reinforced insulation, pollution degree 2
Protection IP40; EN60529
Maximum safe working voltages
Current measurement 600 V ac rms or dc between uninsulated conductor and ground
Voltage measurement 600 V ac rms or dc between either input terminal and ground, or 825 V between energized
phase voltages (delta power config.)
EMC
Emission IEC/EN 61326-1:1997 class B
Immunity IEC/EN 61326-1:1997
Mechanical
Dimensions (length x width x depth) 300 mm x 98 mm x 52 mm (12 in x 3.75 in x 2 in)
Weight including batteries 820 g/1.8 lb
Jaw opening 60 mm
Jaw capacity 58 mm diameter
Cleaning The unit can be cleaned with an Isopropanol impregnated cloth. Do not use abrasives or
other solvents.