Users Manual
30
Additional instrument settings include:
Low pass filter: Switch low pass filter ON or OFF to eliminate high
frequency noise.
PF/DPF Mode: Select either Power Factor or Displacement Power Factor.
PF/DPF Display: Select displayed Power Factor.
Harmonics type: Select either %H1 (fundamental) or %RMS.
Date and Time: For time and date stamping of logged data.
Voltage Range Settings
To select either manual or automatic voltage ranging:
1. Press
JK until Voltage Range is selected.
2. Press
FG to change the setting. The available choices are
AUTO, 4 V, 40 V, 400 V, and 750 V.
3. Press
E to confirm required change.
4. To exit without changing, press
M.
Figure 12 shows the voltage range settings for the Clamp Meter.
eln14.bmp
Figure 12. Voltage Range Settings