35040
Operators Manual
2-10
Table 2-2. Air Density Adjustment Ranges and Resolutions
Units Display Minimum Maximum Standard Resolution
% Error
*
inches of mercury inHg 20.55 31.30 29.90 0.05 0.12
millimeters of mercury mmHg 522 795 760 1.0 0.10
altitude in feet ft -1200 10,000 0 25 0.07
altitude in meters m -360 3050 0 10 0.09
millibars mbar 697 1060 1013 1.0 0.07
hectoPascals hPa 697 1060 1013 1.0 0.07
Units Display Minimum Maximum Standard Resolution
% Error
*
Fahrenheit F 32.0 122.0 71.6 68.0 0.5 0.05
Kelvin K 273.0 323.0 295.2 293.2 0.5 0.09
Celsius C 0.0 50.0 22.0 20.0 0.5 0.09
* The “% Error” column indicates the maximum amount of error that can occur due to the step resolution
of the pressure or temperature units.